12.6
Limiting the mains current consumption
If your mains fusing is below 16 A, the current consumption can be reduced in steps from 16 A to 8 A. The maximum
heating power is reduced correspondingly. Here, take into consideration whether other loads are connected to the same
fused circuit or whether your ECO thermostat is the only load.
Access to the main menu level is obtained by pressing the enter key
.
Selection and confirmation of
Setup
Basic setup
.
The adjacent menu window appears.
As well as
Curr.Consumpt.
the set value is displayed.
Select
Curr.Consumpt.
with .
Change the value with or .
Single figures can be selected by pressing or .
Confirm the input with the enter key .
-
By pressing
(ESC) you are returned to the menu level without any change.
12.7
Entering the offset of the displayed temperature (calibration)
Deviations to the calibrated reference thermometers (e.g. LAUDA DigiCal) can be corrected internally by the "Off-
set" function.
Expert a
pp
endi
x
Expert a
pp
endi
x
Expert a
pp
endi
x
Summary of Contents for ECO GOLD
Page 1: ...Operation manual ECO Gold ...
Page 2: ......
Page 4: ......
Page 81: ...10 2019 i ECO Gold 81 137 W Warnings 55 92 X Xp 106 Xpf 108 ...
Page 82: ...82 137 ECO Gold 10 2019 i ...
Page 83: ...10 2019 i ECO Gold 83 137 ...
Page 84: ...84 137 ECO Gold 10 2019 i ...
Page 85: ...10 2019 i ECO Gold 85 137 Appendix with settings ...
Page 121: ...10 2019 i ECO Gold 121 137 Click on the tab with the mouse and then on the Device manager ...
Page 130: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......