Through flow coolers
16/03/2018 / YAFE0004
7
2
Brief operating instructions
Check through-flow cooler during unpacking for any transport damage, and if necessary, inform the
carrier or the post office.
The unit may NEVER be overturned nor put upside down!
- Set up unit according to chapter 5.
- Connect tubing according to chapter 6.
Thermostats without external systems (
Chapter 6.1).
Thermostats connected to pressure-tight external systems (
Chapter 6.2).
Thermostats with DUPLEX pump or pressure/suction pump connected to
(
Chapter 6.3).
- Secure tubing with clips to prevent slipping-off.
- Check the supply voltage against the data on the rating label. Insert the mains plug.
Switching on:
First switch on thermostat, and then switch on through-flow cooler at
main power switch (green signal lamp lights up).
Switching off:
First switch off through-flow cooler, and then switch off thermostat.
Never allow the through-flow cooler to run without liquid flowing through.