Unpacking and setting up
Never put the device into operation if you have discovered any transport damage.
The unit may NEVER be overturned nor put upside down!
5.1
Unpacking
Carefully packing should prevent transport damage. If, however, the units should arrive damaged, the
carrier or the post office has to be informed so that it can be inspected.
Standard accessories
Cat. No.
2 Screw caps
HKM 032
2 Olives Ø13 mm
HKO 026
1 DLK Mains cable
DLK 10, DLK 25
UK 226
1 valve control line
DLK 45 for Ecoline E 3XX and P control unit thermostat
UK 251
1 power control line
DLK 45 for Ecoline E 3XX and P control unit thermostat
UK 227
1 Operating Instructions
for all DLK
Optional accessories
Cat. No.
1 power control line
DLK 10, DLK 25 for Ecoline E 3XX thermostat and
P control unit thermostat
UK 227
1 Proline control line
DLK 10, DLK 25
UK 263
1 LiBus extension, 5 m
DLK 45 LiBus
EKS 068
1 LiBus T-distributor
DLK 45 LiBus
EKS 073
5.2
Setting up
The condenser of the refrigeration unit is air-cooled. Fresh air is drawn in at the front of the unit and
blown out at the back. Thus, the unit must be set up so that the free air flow is not obstructed.
Recommended distance:
50 centimetre at least.
It is particularly important that the air drawn in is not excessively warm. The unit must not be placed near
a radiator or any other source of heat.
Higher ambient temperatures result in a reduced performance. When the compressor is overloaded
because of high refrigerant pressure or high ambient temperature, the power supply is automatically
interrupted via a temperature switch. The compressor switches on automatically as soon as it has
cooled down.