1
1
.
Frequently Asked Questions
Q1:
The device is on (the screen is measured), but the temperature
cannot be measured
A1:
Confirm whether the product enters the measurement data
viewing mode or parameter setting mode. If you under this mode,
please press the power button (POW key) to shut down (or wait 1
minute to automatically shut down), and then press the temperature
measurement key after rebooting.
Q2:
The measurement temperature is obviously too low
A2:
Confirm whether the device is under the object temperature
measurement mode. If so , press the SET key once to enter the
human body temperature measurement mode. Enter the parameter
setting mode and check whether the temperature compensation
setting is negative. It is recommended to set the value to 0.
Q3:
The measurement temperature is obviously too high
A3:
Enter the parameter setting mode and check whether the
temperature compensation setting is positive. It is recommended to
set the value to 0.
XLE-MD-TM-IR-R1