HVM200 Reference Manual
Measurement Ranges
A-8
A.6 Measurement Ranges
The following table shows the HVM200 dynamic and linearity ranges in root-mean square values. (Peak val-
ues are 1.414 times higher.)
A.7 Frequency Weighting Curves
The following graphs show frequency weighting curves for the HVM200.
Frequency
Weighting
Noise Floor
(typical mV)
Lower Limit
(Under-range
mV)
Lower Limit Lin-
earity Range
(typical mV)
Lower Limit
Linearity Range
(maximum mV)
Upper Limit
(Overload V)
Fa
0.028
0.204
0.092
0.178
5.01
Fb
0.046
0.232
0.133
0.176
5.01
Fc
0.041
0.205
0.116
0.152
5.01
Wb
0.016
0.167
0.073
0.137
5.01
Wc
0.021
0.184
0.089
0.150
5.01
Wd
0.014
0.181
0.072
0.175
5.01
We
0.012
0.193
0.073
0.161
5.01
Wf
0.009
0.185
0.100
0.147
5.01
Wh
0.014
0.087
0.042
0.090
5.01
Wj
0.023
0.167
0.077
0.151
5.01
Wk
0.018
0.144
0.073
0.121
5.01
Wm
0.017
0.103
0.060
0.077
5.01
Summary of Contents for HVM200
Page 2: ...IHVM200 01 Rev C Larson Davis HVM200 Reference Manual...
Page 48: ...HVM200 Reference Manual Frequency Weighting Curves A 9...
Page 49: ...HVM200 Reference Manual Frequency Weighting Curves A 10...
Page 50: ...HVM200 Reference Manual Frequency Weighting Curves A 11...
Page 51: ...HVM200 Reference Manual Frequency Weighting Curves A 12...
Page 63: ...HVM200 Reference Manual Integration Weighting Limits A 24...
Page 64: ...HVM200 Reference Manual Integration Weighting Limits A 25...
Page 65: ...HVM200 Reference Manual Integration Weighting Limits A 26...
Page 66: ...HVM200 Reference Manual Integration Weighting Limits A 27...
Page 67: ...HVM200 Reference Manual Integration Weighting Limits A 28...