HVM200 Reference Manual
Integration Weighting Limits
A-23
A.9 Integration Weighting Limits
All data presented is with a reference sensitivity of 1 mV/m/s
2
. The amplitude values on the tables and fig-
ures scale according to the ratio of selected sensor sensitivity values to 1 mV/m/s
2
.
For the following charts, the valid measurement range for each weighting is shown between the noise floor
and the maximum.
100.0
-39.46
-43.03
-1.45
-3.01
-17.47
-21.05
+1 / -1
125.9
-43.01
-47.51
-3.01
-5.49
-21.04
-25.55
+2 / -2
158.5
-47.46
-52.69
-5.45
-8.66
-25.50
-30.73
+2 / -2
199.5
-52.64
-58.37
-8.64
-12.34
-30.69
-36.42
+2 / -
251.2
-58.27
-64.21
-12.26
-16.22
-36.32
-42.32
+2 / -
316.2
-64.11
-70.24
-16.11
-20.18
-42.16
-48.30
+2 / -
398.1
-70.04
-76.29
-20.04
-24.25
-48.10
-54.40
+2 / -
Single Integration
Double Integration
Weighting
Noise Floor (m/s)
Max (m/s)
Noise Floor (m)
Max (m)
Fa
0.0121
1403.690
0.0150
801.367
Fb
0.0220
1403.690
0.0173
801.367
Fc
0.0003
84.283
0.0003
3.235
Wb
0.0068
563.341
0.0060
317.432
Wc
0.0163
1403.690
0.0143
801.367
Wd
0.0155
1415.048
0.0111
801.367
We
0.0108
1434.733
0.0102
804.139
Wf
0.0802
5740.123
0.1707
8823.712
Wh
0.0003
87.044
0.0002
3.235
Wj
0.0080
695.458
0.0077
397.953
Wk
0.0083
698.669
0.0067
400.711
Wm
0.0049
698.620
0.0034
204.094
Freq
( Hz)
We Ideal
(dB)
We Typ
(dB)
Wj Ideal
(dB)
Wj Typ
(dB)
Wk Ideal
(dB)
Wk Typ
(dB)
Tolerance dB
Summary of Contents for HVM200
Page 2: ...IHVM200 01 Rev C Larson Davis HVM200 Reference Manual...
Page 48: ...HVM200 Reference Manual Frequency Weighting Curves A 9...
Page 49: ...HVM200 Reference Manual Frequency Weighting Curves A 10...
Page 50: ...HVM200 Reference Manual Frequency Weighting Curves A 11...
Page 51: ...HVM200 Reference Manual Frequency Weighting Curves A 12...
Page 63: ...HVM200 Reference Manual Integration Weighting Limits A 24...
Page 64: ...HVM200 Reference Manual Integration Weighting Limits A 25...
Page 65: ...HVM200 Reference Manual Integration Weighting Limits A 26...
Page 66: ...HVM200 Reference Manual Integration Weighting Limits A 27...
Page 67: ...HVM200 Reference Manual Integration Weighting Limits A 28...