Specifications
74
●
Appendix 3
Model
PC210
PC220
pH/ORP/EC/TDS/Sal/Res/Temp (°C/°F)
pH Range
-2.00 to 16.00 pH
Resolution
0.01 pH
Accuracy
±0.01 pH
Calibration Points
USA & NIST (Up to 5), DIN (Up to 6)
pH Buffer Groups
USA, NIST, DIN
ORP Range
±2000 mV
Resolution
0.1 mV (
˂
±1000 mV), 1 mV (
≥
±1000mV)
Accuracy
±0.3 mV (
˂
±1000 mV), 0.3% of reading (
≥
±1000mV)
Calibration Option
Yes
Conductivity Range
µS/cm to 200.0 mS/cm (k=1.0)
Resolution
0.05% full scale
Accuracy
±0.6% full scale, ±1.5% full scale > 18.0 mS/cm
Reference Temperature
15 to 30 °C (adjustable)
Temperature Coefficient
0.00 to 10.00 %/°C
Cell Constants
0.1, 1.0, 10.0
Calibration Points
Up to 4 (Auto) / Up to 5 (Manual)
Units
S/cm, S/m (Auto Ranging)
Total Dissolved Solids (TDS) Range
ppm to 100 ppt (TDS factor=0.5)
Resolution
0.01 ppm (mg/L) / 0.1 ppt (g/L)
Accuracy
±0.1% full scale
TDS Curves
Linear (0.40 to 1.00), EN27888, 442, NaCl
Resistivity Range
0.000
Ω
•cm to 20.0 M
Ω
•cm
Resolution
0.05% full scale
Accuracy
±0.6% full scale,
±1.5% full scale > 1.80 M
Ω
•cm
Salinity Range
0.0 to 100.0 ppt / 0.00 to 10.00 %
Resolution
0.1 ppt / 0.01%
Accuracy
±0.2% full scale
Salinity Curves
NaCl, Seawater
Calibration Option
Yes
Temperature Range
-30.0 to 130.0 °C / -22.0 to 266.0 °F
Resolution
0.1 °C / °F
Accuracy
± 0.5 °C / ± 0.9 °F
Summary of Contents for LAQUA-PC200 Series
Page 1: ......
Page 2: ......
Page 59: ...Data set up 45 EC mode ...
Page 61: ...Data set up 47 EC mode ...
Page 63: ...Data set up 49 EC mode ...
Page 66: ...General set up 52 EC mode ...
Page 68: ...General set up 54 EC mode ...
Page 70: ...General set up 56 EC mode ...
Page 72: ...General set up 58 EC mode ...
Page 75: ...Clock set up 61 EC mode ...
Page 77: ...Clock set up 63 EC mode ...
Page 96: ...Specifications 76 ...
Page 98: ......