Conductivity calibration
20
Calibration
1. After placing the conductivity electrode in the standard solution, press the
key.
2. Meter starts measuring various calibration values with a blinking
on screen.
3. Wait for the
to stabilize (stable calibration reading).
4. Use the
keys to adjust the salinity value.
5. Press the
ENT
key to confirm and save calibration data.
6. Meter displays
DONE
indicating end of the salinity calibration procedure.