Error messages and trouble shooting
69
The response is slow
The indicated value does not change/No response
Cause
How to solve problem
The electrode is dirty.
Clean the electrode.
The electrode is cracked.
Replace the electrode.
The response of electrode is affected
by the sample solution.
It is important to select an electrode that is
appropriate for the sample. Consult your dealer. To
confirm an electrode that is appropriate for the
sample, check the electrode selection guide in our
catalogue, or refer to our website.
Cause
How to solve problem
The electrode is cracked.
Replace the electrode.
The electrode is not connected
correctly.
Connect the electrode correctly.
Keys are locked.
Turn OFF the power, remove the batteries, and then
turn ON the power again.
The instrument is in HOLD state.
Cancel the HOLD state.
Instrument defect
Consult your dealer.