ORP calibration
13
●
ORP/mV Calibration
Calibration is necessary for accurate ORP measurement. To perform ORP calibration, follow
the procedure detailed below:
Prerequisites
• Clean the ORP electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the PC meter and plug in the ORP electrode.
• Prepare standard solution required for calibration.
• Ensure that the meter is in mV measurement mode.
• Dip the ORP electrode into the standard solution ensuring that the solution level is at least
3 cm from the electrode tip.
• Absolute value measurement mode and relative value measurement mode are the two
types of measurement mode available for ORP (mV) measurement.
• In absolute value measurement mode, the handheld meter displays the actual voltage
value.
• In relative value measurement mode, user can adjust the absolute mV value by calibration.
If the mV value is adjusted, the meter automatically indicates relative mV value as
RmV
.
The adjustment mV is applied as an offset to the absolute mV value.
• In the relative mV mode, the absolute mV value can be adjusted by ± 200 mV.
To abort an ongoing calibration process at any point of time, press the
key.
Note
Tip
Summary of Contents for LAQUA-PC200 Series
Page 1: ......
Page 2: ......
Page 59: ...Data set up 45 EC mode ...
Page 61: ...Data set up 47 EC mode ...
Page 63: ...Data set up 49 EC mode ...
Page 66: ...General set up 52 EC mode ...
Page 68: ...General set up 54 EC mode ...
Page 70: ...General set up 56 EC mode ...
Page 72: ...General set up 58 EC mode ...
Page 75: ...Clock set up 61 EC mode ...
Page 77: ...Clock set up 63 EC mode ...
Page 96: ...Specifications 76 ...
Page 98: ......