- 23 -
LANGER
EMV-Technik
DE-01728 Bannewitz
[email protected]
www.langer-emv.com
P1401/P1501 and P1402/P1502 set
EPM 02 / BPM 02 field meters
5.3
BPM 02 dB/dt field meter
Frequency range
500 kHz - 3 GHz
Magnetic flux density
B
calculated based on the voltage
u
out
at the measurement output
Ḃ = K ∙ u
out
K = 122 dB (
T
V ∙ s
)
B = (1 ω
⁄ ) Ḃ
B = Ḃ − 20 log
10
(ω)
EPM 02 dE/dt field meter
Frequency range
500 kHz - 3 GHz
Electric field strength
𝐸
calculated based on the voltage
u
out
at the measurement output
Ė = K ∙ u
out
K = 251 dB (
1
cm ∙ s
)
E = (1 ω
⁄ ) Ė
𝐸 = Ė − 20 𝑙𝑜𝑔
10
(𝜔)
Table 15