33
5
Maintenance
Table 4:Test gas matrix for testing or calibrating the respective probe
Table 5:Test gas composition
Recommendation:
To check the O
2
electrode, use a test gas with 3 vol. % O
2
in N
2
.
5.1.10 Checking/Calibrating with Reference Measurement
•
The system must be in operation (ideally at the desired operating point).
•
Perform the calibration as described in Chapters
5.1.2 Checking/Calibrating the O
and
5.1.2 Checking/Calibrating the O
5.1.11
Consumables
Lambda Probe
Average lifetime
–
natural gas and light fuel oil: 3 ... 5 years
–
with heavy fuel oil, coal and biogas: 1 ... 3 years
Functional check/calibration
with test gas
Probe
LS2
1
Offset
A
2
O
2
B
Test gas
Composition
O
2
[Vol.%]
N
2
[Vol.%]
A
21
Rest
B
3
Rest