Care and inspection work 9
Issue EN 2022/05 PUK 6 Operating manual
49/56
9
Care and inspection work
The device must always be switched off before care and inspection work.
9.1
Care and inspection schedule
Interval
Care and inspection work
Comments
Daily
Check working environment
Clean if necessary
Check the condition and cleanliness of the de-
vice
Clean if necessary
Check the gas hose for leaks
As required
Clean the screw connection of the electrode in
the handpiece to ensure electrical contact.
The electrode can be reground.
Replace electrode when worn out.
Tab. 9.1 Care and inspection schedule
9.2
Carry out care and inspection work
Cleaning the PUK 6
•
Use a dry / slightly damp cloth to remove dust and carefully clean the
surfaces of the device.
•
A non-abrasive glass cleaner (e.g. spectacles cleaner) can be used for
the screen on the front of the display.
Regrinding electrodes
1)
Switch off the device.
2)
Remove the electrode from the handpiece.
3)
Use the electrode grinding motor with a fine or medium-fine grinding
wheel to grind the electrode.
4)
Grind the electrode at a 15° angle. With the grinding motor from the
PUK 6 original accessories, this angle is predetermined by the guide
holes.
Now the electrode can be reinserted into the handpiece.