9
OPERATIONAL MATRIX
Function /
Resulting Action
Power
Status
Mode
Setting
Required Button Press Sequence
ON/OFF
Any
Any
Momentary press of the
ON/OFF
button
Water Saturated
Air Calibration
On
Any
Place electrode in electrode cap. Press &
hold
CAL
button for 2 seconds.
Zero Calibration On
Any
Place electrode in zero DO solution. Wait
for stability. Press & hold
CAL
button
for 2 seconds. Also works with probe
removed. (Cal at zero current).
Store Reading On
Any
Momentary press of
MODE
button.
Stores and holds reading. ‘HOLD’
displayed.
Hold Release
On
While In
Hold Mode
Momentary press of the
MODE
button.
Enter Memory
Retrieval
On
Any
Momentary press of
CAL
button followed
by a momentary press of the
MODE
button within 4 seconds. If no data is
stored in memory, ‘End’ is displayed
briefly, and then returns to last mode.
Scroll Stored
Readings
On
Memory
Recall
Momentary press of
MODE
button.
Exit Memory
Retrieval
On
Memory
Recall
Momentary press of
CAL
button.
Clear Stored
Memory
On
Any
Measure-
ment Mode
Press and hold
ON/OFF
button for 4
seconds, ‘clr’ is displayed.
Change
Measurement
Mode
On
Any
Press and hold
MODE
button for at least
two seconds. Modes will scroll until button
is released.
Enter Salinity
Compensation
On
Any
Press and release the
CAL
button twice
quickly in succession. ‘SAL’ is displayed.
Changing
Salinity
Compensation
On
SAL
Momentary press of
MODE
button. Each
button press increases the ratio by 1 ppt
(part per thousand). The value cycles from
0 to 50 ppt.
Exit Salinity
Compensation
On
SAL
Press and release
CAL
button for 2
seconds to enter Altitude Compensation.
Or press and release
CAL
button one
more time to enter measurement mode. It
is necessary to press
CAL
button to save
changes. If the unit times out, no changes
are saved.