background image

8

2. Slowly stir the sample with the electrode and wait for the reading to 

stabilize. This may take a while, depending on the electrode history.

3. Press 

CAL

 until ‘CAL’ is shown on the display. The readings will blink ‘101.7’ 

and ‘SA’ will appear. ‘SA’ will not appear if the calibration fails.

4. When the calibration is complete ‘End’ will appear and the meter will return 

to the measurement mode. NOTE: Sodium Sulfite can become deposited 

on the electrode and on the coined surface of the electrode retaining collar. 

The presence of the Sodium Sulfite will negatively affect dissolved oxygen 

measurements unless it is removed completelyfrom the electrode.

DISSOLVED OXYGEN MEASUREMENT

(Be sure the electrode is filled and fully polarized. Calibrate daily.)
1. Cover the electrode with the electrode cap. The sponge in the electrode cap 

should be moistened (not soaked) with distilled water or tap water. 

2. Press 

ON/OFF

 to turn the meter ON. The meter will self calibrate. 

3. Press and hold 

MODE

 to select the desired unit of measure. 

4. Remove the electrode cap. 
5. Place the electrode in the sample to be measured.  Stir the sample with the 

electrode.

6. Wait for the reading to stabilize.  
7. Press 

ON/OFF

 to turn the meter OFF. Rinse the electrode with distilled 

water or tap water. Replace the electrode cap. 

NOTE:  The unit of measure selected when the meter was last turned OFF will 

be retained. Stir the sample with the electrode to remove any air bubbles that 

may be trapped on the membrane surface and to obtain accurate results. Do 

not submerse the electrode above the bottom edge of the electrode collar. 

For maximum accuracy, allow sufficient time for the temperature of the probe 

to reach the temperature of the sample before taking a reading. The larger 

the difference in temperature between the electrode and the sample the 

longer it will take for the reading to stabilize. Stabilization will be indicated by 

a stable temperature reading on the display. Stabilization time can vary from 

10 seconds to 5 minutes. The DO Tracer uses a polarographic sensor, which 

consumes oxygen at the sensing surface. This requires a constant movement 

of the sample across the membrane to maintain a constant dissolved oxygen 

level. It is recommended that the sample be stirred gently with the electrode as 

the measurement is made, or if in a laboratory environment, the sample should 

be stirred using a stir plate.

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for Tracer PockeTester

Page 1: ...brate daily Polarize weekly Dissolved Oxygen Measurement Procedure 1 Press ON OFF button to turn meter ON 2 Place TRACER in sample Stir 3 Wait for reading to stabilize 4 Record result 5 Press ON OFF to turn meter off 6 Rinse electrode 7 Replace electrode cap WARNING This set contains chemicals that may be harmful if misused Read cautions on individual containers care fully Not to be used by childr...

Page 2: ... 36 x 173 x 41mm 1 4 x 6 8 x 1 6 110g 3 8 oz Measurement Range Resolution Accuracy Saturation 0 to 200 0 0 1 2 0 FS full scale Dissolved Oxygen 0 to 20 00 mg l 0 01 mg l 2 FS Concentration 0 to 20 00 ppm 0 01 ppm 2 FS Temperature 0 to 50 C 0 1 C 1 0 C 32 to 122 F 0 1 F 0 to 99ºF 1 0 F 100 F 1 8 F CONTENTS Dissolved Oxygen TRACER Kit Code 1761 Includes DO TRACER Body DO Bonded Membrane Cap Assembly...

Page 3: ...ng the membrane cap Powering the Tracer The Tracer uses four 4 CR2032 Lithium Ion Batteries Press the ON OFF button to turn the meter on or off If the batteries are weak the BAT indicator will appear on the display The auto power off feature shuts the Tracer off automatically after approximately 10 minutes of inactivity The auto power off feature may be temporarily disabled for convenience or for ...

Page 4: ...momentarily press ON OFF to turn the Tracer ON 3 Release CAL SELF CAL will be displayed 4 Repeat Steps 1 3 to change to the other temperature unit Changing the Salinity Compensation 1 With the Tracer ON momentarily press CAL twice in quick succession SAL will be shown on the lower part of the display 2 Momentary press MODE Each press of MODE increases the salinity compensation by 1ppt part per tho...

Page 5: ...eter off and on again using ON OFF Low Battery Indicator When the battery voltage falls below the operating threshold BAT will appear on the display Refer to the Maintenance section for battery replacement information Storing Readings 1 Press MODE to store a reading The storage location number will be shown on the lower display and the main display will show the stored reading The meter will enter...

Page 6: ...e cathode If a membrane cap is removed it can not be reinstalled because the membrane will no longer stretch properly over the cathode 4 Rinse the cathode and anode with deionized water 5 Set the membrane cap assembly on a flat surface 6 Fill the membrane cap assembly with Electrolyte Solution DO600 EL to the bottom of the threads on the inside of the cap To remove air bubbles hold the cap steady ...

Page 7: ...orner of the display when the meter is OFF If the asterisk is absent turn the meter ON and allow the electrode to fully polarize This may take 2 3 minutes The asterisk will appear to indicate full polarization Calibration Calibration should be performed daily Be sure the electrode is filled and fully polarized before calibrating 1 Turn the meter ON 2 Press and hold MODE until is displayed 3 Remove...

Page 8: ...ith the electrode 6 Wait for the reading to stabilize 7 Press ON OFF to turn the meter OFF Rinse the electrode with distilled water or tap water Replace the electrode cap NOTE The unit of measure selected when the meter was last turned OFF will be retained Stir the sample with the electrode to remove any air bubbles that may be trapped on the membrane surface and to obtain accurate results Do not ...

Page 9: ...urns to last mode Scroll Stored Readings On Memory Recall Momentary press of MODE button Exit Memory Retrieval On Memory Recall Momentary press of CAL button Clear Stored Memory On Any Measure ment Mode Press and hold ON OFF button for 4 seconds clr is displayed Change Measurement Mode On Any Press and hold MODE button for at least two seconds Modes will scroll until button is released Enter Salin...

Page 10: ...unit times out no changes are saved Override Auto Power Off On Any Press CAL button momentarily then simultaneously press and hold MODE and ON OFF buttons for 2 seconds Default Reset Off OFF mode Simultaneously press ON OFF CAL and MODE buttons momentarily dFLt will be displayed TROUBLESHOOTING If the display appears to be frozen it is possible that the Data Hold mode hasbeen inadvertently activat...

Page 11: ...D mode Meter is locked Release HOLD by momentarilly pressing MODE button Remove batteries Press ON OFF button Replace batteries and press ON OFF button Stored readings will be lost MAINTENANCE Battery Replacement 1 Twist off the battery compartment cover 2 Holding the battery housing in place with a finger pull out the battery carrier using the two small tabs 3 Replace the four 4 CR2032 batteries ...

Page 12: ...electrode the membrane is stretched by the cathode and cannot be reused It must be replaced See page 6 Electrode Storage Store the TRACER with the Electrode Storage Cap in place The sponge in the cap should always be moist without excess liquid Electrode Cleaning Recommendations Contaminant Cleaning Solution Procedure Salt build up Deionized water Soak in 20 or weaker Acetic Acid solution for 1 mi...

Reviews: