5. Rinse a clean tube (0290) with the
chlorine standard. Fill the tube to
the 10 mL line with the chlorine
standard. Dry tube with a lint-free
cloth. Put on a dry positioning ring.
Cap the tube.
6. Open the meter lid. Insert the tube
into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter. Close
the lid and press
*
I
OK
to select
Scan Blank
.
F r e e C h l o r i n e
( L )
* S c a n B l a n k
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
7. Remove the tube from the meter.
Add 5 drops of DPD 1A Free
Chlorine Reagent (P-6740) and
mix.
8. Add 5 drops of *DPD 1B Free
Chlorine Reagent (P-6741). Cap
and mix. Read within 30 seconds.
CALIBRATION & ANALYSIS • CHLORINE
65
DPD
1B
DPD
1A