4. Press the
*
I
OK
to select
Units
.
T u r b i d i t y
* U n i t s
C a l i b r a t i o n
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
5. Scroll down and then press
*
I
OK
to select units.
Available units are:
NTU
(Nephelometric Turbidity Units);
FNU
(Formazin Nephelometric
Units);
ASBC
(American Society of
Brewing Chemists);
EBC
(European
Brewery Convention)
U n i t s
* N T U
F N U
A S B C
E B C
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
Note:
If Attenuation is chosen as a calibration curve. The result will be reported in
FAU (Formazin Attenuation Units).
6. Press
to exit to a previous menu or
make another menu selection or
press
OFF
to turn the meter off.
SELECTING A TURBIDITY CALIBRATION CURVE
1. Press
ON
to turn the meter on.
1 . 3
2. Scroll down and then press
*
I
OK
to
select
Options
.
M a i n M e n u
M e a s u r e
D a t e
* O p t i o n s
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
26
OPTIONS & SETUP • TURBIDITY