6. Remove the tube from the meter.
Empty the tube. Rinse the tube with
the color standard. Fill the tube to
the 10 mL line with the color
standard.
7. Open the meter lid. Insert the tube
with the color standard. Align the
index notch on the positioning ring
with the index arrow on the meter.
Close the lid. Press
*
I
OK
to select
Scan Sample
.
C o l o r
* S c a n S a m p l e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
8. View the result.
C o l o r
2 8 . 0
c u
* S c a n S a m p l e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
9. Press
and then press
*
I
OK
to
select
Calibrate
.
C o l o r
2 8 . 0
c u
* C a l i b r a t e
~
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
10. Use
or
to change the
highlighted digits on the display to
match the concentration of the color
standard. Press
*
I
OK
to accept a
digit and move to the next digit.
C a l i b r a t e
0 2 8 . 0
q
,
Q
,
p
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
C a l i b r a t e
0 2 8 . 0
q
,
Q
,
p
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
72
CALIBRATION & ANALYSIS • COLOR
2
0