Lambrecht ARCO-Modbus 14581 Manual Download Page 5

Manual 

Combined Wind Sensor ARCO-Modbus

(14581) ARCO-Modbus

Manual

2.8.2  Power and Signal Connection

If the sensor is mounted in the correct manner and connected 

with the right cable (accessory), you can attach the wires for 

the power supply and signal outputs to the data acquisition 

equipment (computer).
The ARCO sensor requires a 4-pole M12 plug connector. 

The cable shield should be connected with both ends at the 

ground wire (PE).

To reduce the risk of inductive interference the sen-

sor must be properly grounded (screening on both 

sides).

The external connection is carried out via a central connec-

tor which is located in the housing base. For further details 

about  the electrical connection please see section "Electrical 

connections“.
The typical power supply of the sensors is 24 VDC with a 

current consumption of max. 920 mA (incl. heating). The 

input voltage range can be 20...28 VDC. The heating of the 

ARCO-Modbus has a heating power of 2·9 W.
Under most climatological conditions the heating prevents 

blocking of the moving sensor parts. Neither the cup rotor nor 

the wind vane are heated. In the case of icing or formation 

of ice on the moving sensor element the function is restricted 

for the period of icing.
The output signal of the sensor conforms to the Modbus RTU 

to RS485 standard.

2.8.3  Safety Regulations

Because the wind sensor is often mounted in ex-

posed locations at dangerous heights the installation 

personnel have to pay attention to the relevant safety 

regulations. During the electrical installation work the 

external circuitbreaker must be switched off.
It is not permitted for unauthorized persons to open 

the housing!

3 Maintenance

3.1 

Regular Maintenance and Calibration

The sensors are very low-maintenance and designed for a 

very long lifetime. Regular visual checks are recommended, 

regarding dirt build-up on the surface caused by the weather. 

Whenever this is the case, we strongly recommend  that you 

clean it. A regular visual and functional check of the wind 

sensor is recommended.

If reference measurements are necessary, it must be 

stringently noted that a comparison of the measured 

values is given only if the measurements take place 

under same conditions. I.e. the reference equipment 

must be used very close to the sensor!

The sensor is a measuring instrument and thus user specific 

standards apply regarding the period of recalibration. 
Manufacturer's recommendation: 2 years.
For the long-term assurance of the function and accuracy of 

the proposed components, we are pleased to offer you our 

professional maintenance and calibration services.

3.2 

Visual Check and Cleaning

The use of the sensor under the respective environmental 

conditions requires certain actions. It is thus recommendable 

to clean the outside of the housing and shelter within specific 

intervals. The intervals are dependant on the environmental 

conditions and the degree of soiling. We recommend a regular 

visual and functional check.

In case you should be faced with any specific problems, which 

you are unable to solve, please contact the LAMBRECHT 

meteo service:
Tel.:  

+49-(0)551-4958-0

Fax:  

+49-(0)551-4958-312

E-Mail:   [email protected]

4 Transport

In case it is necessary for you to ship or transport the sensor, it 

must be carefully packed to prevent damages during transport.

5

Summary of Contents for ARCO-Modbus 14581

Page 1: ...g value of 0 3 m s Very large measuring range from 0 3 to 75 m s Quick and easy installation by pipe mounting and M12 plug connector Integrated regulated heater for optimal heating of the sensor heads...

Page 2: ...uipment Content 1 Introduction 3 2 Start Up 3 2 1 Installation Conditions 3 2 2 Tools and Installation Aids 3 2 3 Unpacking the Sensor 3 2 4 Goods Inspection 3 2 5 Power Supply 3 2 6 Mounting the Cup...

Page 3: ...luminum alloys The housing the cup rotor and the wind vane are anodized An electronically controlled sensor heating device allows for operation of the sensor over the wide range of 30 70 C 2 Start Up...

Page 4: ...d direction measurements the north mark on the sensor must be aligned with the geographical north direction To adjust the wind sensor in a firm and correct manner into the north direction this item is...

Page 5: ...cuitbreaker must be switched off It is not permitted for unauthorized persons to open the housing 3 Maintenance 3 1 Regular Maintenance and Calibration The sensors are very low maintenance and designe...

Page 6: ...Manual Combined Wind Sensor ARCO Modbus 14581 ARCO Modbus Manual 5 Dimensional Drawing and Electrical Connection ARCO Modbus 6...

Page 7: ...Manual Combined Wind Sensor ARCO Modbus 14581 ARCO Modbus Manual Electrical connection ARCO Modbus 7...

Page 8: ...e by the divisor Values of 9999 indicate an internal sensor error 6 3 Standard Configuration Default Baud rate 19200 Baud Byte frame 8E1 1 start bit 8 data bits 1 parity bit even parity 1 stop bit RTU...

Page 9: ...ss type 30002 Wind speed average m s 10 1 Read only 30003 Wind speed maximum m s 10 1 Read only 30004 Wind speed minimum m s 10 1 Read only 30202 Wind direction average 10 1 Read only 30203 Wind direc...

Page 10: ...d Input Register 4 Func Desk Byte count 2 Data 00 14 Checksum OK 3FB9 6 7 Descriptive Sensor Parameter Registers Holding Register Register Parameter name Quantity of registers Remark Access type 40050...

Page 11: ...01 06 48 05 10 9C 41 00 01 7E 09 LEN 9 Transmission Query Source Master Dest Slave 5 Function Write Multiple Register 16 Func Desk Address 40001 Quantity 1 Byte count 2 Register values 00 01 Checksum...

Page 12: ...ely Sensor cable 15 m 4 pole M12 plug Id No 32 14567 060010 various lengths possible Options 9340 Visualisation and evaluation software MeteoWare CS Basic Id No 36 09340 000000 Standards EMC EMI EN 60...

Reviews: