Kiosk placement
Do not place the device directly :
- Facing the sun or in a location where the sun is directly facing the person
- In front or near an air outlet of an air conditioning or radiator device during measurement
- In front or near to strong infrared emitting light
- In a room where heating/cooling ducts are in the ceiling or directly below ceiling ducts
The above can cause changes to the forehead temperature. Measurements should be
taken in a stable environment.
Do not measure body temperature in an environment with:
- Strong electromagnetic interference (examples include places close to a working
microwave, induction appliances) as EM interference may cause errors in the reading
or even device failure.