Lake Shore Model 480 Fluxmeter User’s Manual
Computer Interface Operation
6-33
PEAK?
Query Peak Hold Function Off/On.
Input:
PEAK?
Returned:
0
or
1
. Format:
n[term]
.
Remarks:
Queries the peak hold function.
0
= Normal reading,
1
= Peak hold.
PEAKM
Configure Peak Hold Mode.
Input:
PEAKM <mode>
Returned:
Nothing.
Remarks:
Configures the peak hold mode.
0
= Positive peak,
1
= Negative peak,
2
= Both positive and
negative peaks. Capturing both peaks decreases sampling frequency by a factor of 4.
PEAKM?
Query Peak Hold Mode.
Input:
PEAKM?
Returned:
0
,
1
, or
2
. Format:
n[term]
.
Remarks:
Queries the peak hold mode.
0
= Positive peak,
1
= Negative peak,
2
= Both positive and
negative peaks. Capturing both peaks decreases sampling frequency by a factor of 4.
PKNEG?
Query Negative Peak Reading.
Input:
PKNEG?
Returned:
±nn.nnnE±nn
Remarks:
Returns current negative peak reading in exponential form in the currently selected units.
PKPOS?
Query Positive Peak Reading.
Input:
PKPOS?
Returned:
±nn.nnnE±nn
Remarks:
Returns current positive peak reading in exponential form in the currently selected units.
PKRST
Initiate Peak Reset Command.
Input:
PKRST
Returned:
Nothing.
Remarks:
Resets the peak hold circuit and the peak hold values.
PROBE
Set Probe Serial Number and Date.
Input:
PROBE <serial number>,<date>
Returned:
Nothing.
Remarks:
Sets the serial number and date for the attached probe. A probe with a PROM attachment
(Lake Shore part #FCBL-6) is required for this function. Enter up to 10 characters for a serial
number and enter a date in the form
mmddyyyy
(month, day, year). This command does not
save the information to the probe itself. The COILSAVE command must be used to save the
serial number, date, and coil parameters to the probe.
Example:
PROBE User 5,12291998
sets the serial number to User 5 and the date to 12/29/1998.
PROBE?
Query Probe Serial Number and Date.
Input:
PROBE?
Returned:
<serial number>,<date>
. Format:
aaaaaaaaaa,nnnnnnnn
.
Remarks:
Returns the serial number and date for the probe attached during power up. If a new probe is
attached, cycle the power to load new probe data. A probe with a PROM attachment (Lake
Shore part #FCBL-6) or a Lake Shore probe is required for this function. Returns 10
characters for a serial number and returns a date in the form
mmddyyyy
(month, day, year).