Lake Shore Model 480 Fluxmeter User’s Manual
8-14
Service and Calibration
8.9.7.2
Corrected Output Calibration
1. Connect DVM to Model 480 Corrected Output, set to DC.
2. Set Model 480 Analog output to manual, -100%.
3. Corrected Offset Constant = –(Voltmeter reading 1 + 10.000 Vdc) / 20.
4. Set Model 480 Analog output to manual, +100%.
5. Corrected DAC Gain Constant = 20 / ((Voltmeter reading 2) – (Voltmeter reading 1)).
6. Send Corrected DAC Offset Constant using the form “CALCOFF .xxxxxx”.
NOTE:
This value must be between –0.005 and 0. A value outside this range indicates a major
malfunction of the Model 480 that requires repair.
7. Send Corrected DAC Gain Constant using the form “CALCGAIN <.xxxxxx>“.
NOTE:
This value must be between 0.95 and 1.0. A value outside this range indicates a
major malfunction of the Model 480 that requires repair.
8.9.8 Finalize Calibration
1. Finalize calibration by sending “POKEM F207,31” over the computer interface.
2. Set the unit serial number using the form “SNUM xxxxxx”, sent over the interface.
3. The calibration is now complete.