Lake Shore Model 480 Fluxmeter User’s Manual
6-32
Computer Interface Operation
LOCK?
Query Front Panel Keypad Lock Function.
Input:
LOCK?
Returned:
0
or
1
. Format:
n[term]
.
Remarks:
Queries the front panel keypad lock function.
0
= Unlocked,
1
= Locked.
MODE
Configure Remote Interface Mode.
Input:
MODE <mode>
.
Returned:
Nothing.
Remarks:
Configures the remote interface mode.
<mode>
specifies which mode to operate.
0
= local,
1
= remote,
2
= remote with local lockout. Press the front panel
Local
key to set the Model
480 to local provided the key has not been disabled by local lockout. The Model 480 powers
up in local mode.
MODE?
Query Remote Interface Mode.
Input:
MODE?
Returned:
<mode>
. Format:
n[term]
.
Remarks:
Returns the remote interface mode.
0
= local,
1
= remote,
2
= remote with local lockout.
OPTIN?
Query Optional Input.
Input:
OPTIN?
Returned:
0
or
1
. Format:
n[term]
.
Remarks:
Queries the optional input on the rear panel.
0
= open or logic high,
1
= shorted or logic low.
PCTMUL
Set Percent Multiplier Constant.
Input:
PCTMUL
nnn.nnnE
nn
Returned:
Nothing.
Remarks:
Sets the percent multiplier used to calculate units of percent. Enter up to 6 digits and a
decimal point in exponential form.
PCTMUL?
Query Percent Multiplier Constant.
Input:
PCTMUL?
Returned:
nnn.nnnE
nn
Remarks:
Returns the percent multiplier constant used to calculate units of percent. Returns up to
6 digits and a decimal point in exponential form.
PCTSET
Initiate Set Percent Command.
Input:
PCTSET
nnn.nnnE
nn
Returned:
Nothing.
Remarks:
Recalculates a percent multiplier for the currently measured field. The number part of the
command is the desired reading value of the current field. The coil must be in the field when
the command is issued.
Example:
PCTSET 50.00[term]
: Equates a display of 50% with the current reading. Anytime the
reading equals the current reading, a value of 50% displays.
PEAK
Configure Peak Hold Function Off/On.
Input:
PEAK <off/on>
Returned:
Nothing.
Remarks:
Configures the peak hold function.
0
= Normal reading,
1
= Peak hold.