Lake Shore Model 480 Fluxmeter User’s Manual
Computer Interface Operation
6-29
COILR?
Query Coil Resistance.
Input:
COILR?
Returned:
nnn.nnnE
nn
Remarks:
Returns coil resistance for current coil in units of ohms (
). This is the DC resistance of the
coil itself. Returns up to 6 digits and a decimal point in exponential form.
COILSAVE
Initiate Coil Save Command.
Input:
COILSAVE <coil number>
Returned:
Nothing.
Remarks:
Saves the current coil parameters to a coil location.
1 - 10
= internally stored coil parameters,
11
= probe data. Any previous information in the coil location will be overwritten. A probe with
a PROM attachment (Lake Shore Part Number FCBL-6) is required to write to probe data
(coil number 11).
DACCRS
Set Coarse Drift Adjustment DAC Value.
Input:
DACCRS
nnn.nnnE
nn
Returned:
Nothing.
Remarks:
Sets the percentage of full scale of the coarse drift adjustment DAC. Enter up to 6 digits and
a decimal point in exponential form. Valid values are from
+100%
to
-100%
.
DACCRS?
Query Coarse Drift Adjustment DAC Value.
Input:
DACCRS?
Returned:
nnn.nnnE
nn
Remarks:
Returns the percentage of full scale of the coarse drift adjustment DAC, up to 6 digits and a
decimal point in exponential form.
DACFINE
Set Fine Drift Adjustment DAC Value.
Input:
DACFINE
nnn.nnnE
nn
Returned:
Nothing.
Remarks:
Sets the percentage of full scale of the fine drift adjustment DAC. Enter up to 6 digits and a
decimal point in exponential form. Valid values are from
+100%
to
-100%
.
DACFINE?
Query Fine Drift Adjustment DAC Value.
Input:
DACFINE?
Returned:
nnn.nnnE
nn
Remarks:
Returns the percentage of full scale of the fine drift adjustment DAC, up to 6 digits and a
decimal point in exponential form.
DCRES
Set DC Resolution High/Low
Input:
DCRES
Returned:
Nothing.
Remarks:
Sets the DC Resolution of the unit, where 0 = 5¾ digits and 1 = 4¾ digits.
DCRES?
Query DC Resolution High/Low
Input:
DCRES?
Returned:
0
, or
1
. Format:
n[term]
.
Remarks:
Returns the DC Resolution of the unit, where 0 = 5¾ digits and 1 = 4¾ digits.