Lake Shore Model 460 Gaussmeter User’s Manual
4-40
Remote Operation
Probe Specific Commands (Continued)
ONOFF?
Probe On/Off Query
Input:
ONOFF?[term]
Returned:
<state>[term]
Format:
n
(Refer to command for description)
SNUM?
Probe Serial Number Query
Input:
SNUM?[term]
Returned:
<serial>[term]
Format:
annnnnnnnn
<serial>
The current probe serial number format is Hnnnnn, though there is room for
up to a ten-character response.
TCOMP
Temperature Compensation Command
Input:
TCOMP <state>[term]
Format:
n
<state>
0 = Off, 1 = On
Remarks:
Turns set temperature compensation On or Off. If off, probe temperature compensation, if
present, is ignored. Refer to Paragraph 3.7.
TCOMP?
Temperature Compensation Query
Input:
TCOMP?[term]
Returned:
<state>[term]
Format:
n
(Refer to command for description)
TYPE?
Probe Type Query
Input:
TYPE?[term]
Returned:
<type>[term]
Format:
n
<type>
0 = High Sensitivity (HSE)
1 = High Stability (HST)
2 = Ultra-High Sensitivity (UHS)
Remarks:
Refer to Paragraph 5.2.
ZCAL
Zero Probe Command
Input:
ZCAL[term]
Remarks:
This command initiates the Zero Probe function. Place probe in the Zero Gauss Chamber first
and then enter the ZCAL command. Refer to Paragraph 3.3.