Lake Shore Model 460 Gaussmeter User’s Manual
Operation
3-19
Corrected Analog Out (Continued)
Enter the numbers 1.5 on the numerical keypad and press the
Enter
key. A maximum output of
+1.5 kG has now been placed into memory. Upon pressing
Enter
, the following display will appear.
En t e r Mi n o u t p u t
- 1 . 5 0 0 0 k G
Enter the numbers –1.5 on the numerical keypad and press the
Enter
key. A minimum output
of –1.5 kG has now been placed into memory. Changes to the Corrected Analog Output are
immediately observable.
The second example is an asymmetrical scaling which demonstrates the versatility of user selectable
scaling.
0 V
+1 V
+2 V
+3 V
–3 V
–2 V
–1 V
+1.5 kG
+2 kG
+2.5 kG
+3 kG
0 kG
+0.5 kG
+1 kG
Display
Reading
Output
Voltage
To enter this scale, press the
Analog Out
Key. Press the
Analog Out
,
s
, or
t
key to cycle the arrow
(
→
) to User as shown below.
Se l e c t Wi t h ° ®
An a l o g Ou t ¡ De f
¡ Us e r
Press the
Analog Out
,
s
, or
t
key to cycle the arrow (
→
) to Def (Default). Press the
Enter
key. The
display will automatically step to the Analog Output Source selection display shown below:
Se l e c t Wi t h ° ®
An a l o g Ou t So u r c e X
Press the
s
or
t
key to cycle the analog output source from channel X, Y, Z, or Vector. In this case,
we chose Channel X. Press the
Enter
key and observe the following display.
En t e r Ma x o u t p u t
+3 . 0 0 0 0 k G