Lake Shore Model 460 Gaussmeter User’s Manual
Remote Operation
4-39
Device Specific Commands (Continued)
UNIT
Gauss/Tesla Units Command
Input:
UNIT <unit>[term]
Format:
a
<unit>
G = gauss, T = tesla. Refer to Paragraph 3.8.
UNIT?
Gauss/Tesla Unit Query
Input:
UNIT?[term]
Returned:
<unit>[term]
Format:
a
(Refer to command for description)
VSRC
Vector Magnitude Source Command
Input:
VSRC <source>[term]
Format:
n
<source>
1 = XYZ, 2 = XY, 3 = XZ, 4 = YZ, 5 = X – Y
Remarks:
Sets the source of Vector Magnitude channel calculations. Refer to Paragraph 3.1.5.
VSRC?
Vector Magnitude Source Query
Input:
VSRC?[term]
Returned:
<source>[term]
Format:
n
(Refer to command for description)
4.3.4
Probe Specific Commands
FCOMP
Field Compensation Command
Input:
FCOMP <state>[term]
Format:
n
<state>
0 = Off, 1 = On
Remarks:
Turns set field compensation On or Off. If Off, probe field compensation table, if present,
is ignored. Refer to Paragraph 3.7.
FCOMP?
Field Compensation Query
Input:
FCOMP?[term]
Returned:
<state>[term]
Format:
n
(Refer to command for description)
ONOFF
Probe On/Off Command
Input:
ONOFF <state>[term]
Format:
n
<state>
0 = Off, 1 = On
Remarks:
Sets probe on/off status. First specify channel X, Y, or Z with the CHNL command.