Model 240 Quick Start Guide

Quick Start Guide

240 Series Cryogenic Sensor Input Module

Summary of Contents for 240

Page 1: ...1 https www lakeshore com 240 Model 240 Quick Start Guide Quick Start Guide 240 Series Cryogenic Sensor Input Module...

Page 2: ...rotected against harmful ingress of water Ground the instrument To minimize shock hazard the instrument is equipped with a grounded connection to the DIN rail Connect the DIN rail to an electrical gro...

Page 3: ...ormation for getting started with your input modules For further information please see the Model 240 user s manual Front mounted OLED screen Cryogenic temperature sensor inputs 5 pin connectors Power...

Page 4: ...slightly with shipping companies Keep all damaged shipping materials and contents until instructed to either return or discard them 4 Open the shipping container and keep the container and shipping ma...

Page 5: ...oltage and current leads should then be covered with a braided or foil shield that is connected to the shield pin of the instrument This type of cable is available through local electronics suppliers...

Page 6: ...e USB disconnect it from the computer before beginning PLC communication Physical Installation The 240 Series input module has been specifically designed to mount on a DIN rail To minimize shock hazar...

Page 7: ...his is just the beginning of what is possible with the Model 240 For additional features please see the Model 240 user s manual Basic operation 1 Connect a power adapter to the input module 2 Connect...

Page 8: ...ck Start Guide 6 Turn Sensor Excitation ON Turn Sensor Excitation off to disable any unused inputs 7 Load the sensor calibration file under Sensor setup that matches the sensor by navigating to the ap...

Page 9: ...acy or High speed NOTE The Current Reversal option is available if a resistive sensor is connected and the High accuracy mode is selected NOTE High speed ID is only available for the Model 240 2P 10 U...

Page 10: ...exactly with the markings on the dial Please do not use the sensor simulator as a method for verifying accuracy of the 240 Series input module The evaluation kit includes two unique accessories for t...

Page 11: ...the terminal connector when connecting and disconnecting to the 240 Series input module 3 Plug the terminal connector into the desired input of the 240 Series input module The switch should face towar...

Page 12: ...ort The Lake Shore Forum is also a great place to look for solutions to post issues and to share successes https forums lakeshore com For further documentation and information please see https www lak...