background image

User’s Manual

Fax: 765-423-4111 . www.lafayetteinstrument.com . [email protected]

7

9.  FR RT STIM LITE – same as #8 for the Front Right Stimulus Light.
10.  Auxiliary Output #3 – Press the Up button and Enter to turn a device, e.g. house light, that is connected to 

the auxiliary output #3 on the rear panel ON during magazine training.  Press Enter only to leave the Aux 

#3 output disabled.

11.  Auxiliary Output #4 – same as #10 except for Aux #4 output.
12.  Press Run to Begin – Counters are displayed to monitor animal responses, shaping responses and rewards.  

See Schedule Descriptions for additional details.

Schedule Descriptions

The step-by-step setups of each schedule are similar to that described for Magazine Training and will not 

be repeated in this section except for settings unique to the operation of the schedule.  Use the Down/Up 

buttons  to  scroll  down  or  up  the  list  of  available  schedules.    A  separate  extinction  schedule  is  no  longer 

provided.  Simply use Magazine Training, Partial Reinforcement, or Fixed Ratio without a reinforcement device 

selected.  When using Partial Reinforcement, program the stimulus light to come on when the shaping button 

is pressed to look at spontaneous recovery.

Magazine Training

a.  Both front press bars and the shaping push button increment their respective counters.
b.  Reinforcement is given for each selected/acceptable response and for each push button activation.
c.  The reinforcement counter is incremented for all reinforcements.
d.  The CS lights are available and turn ON at the intensity selected from 1 - 20 when the Run command is 

issued.  The CS remains ON for as long as the schedule is running. 

e.  A single Reset pauses the schedule and turns selected outputs OFF.  The green status light flashes while 

the schedule is paused.  Press Run to continue and Reset to terminate the schedule.

f.  The Tones and Shock are OFF.  

Partial Reinforcement (PRF)

a.  All  press  bars,  the  Omni-directional  lever,  and  the  shaping  push  button  increment  their  respective 

counters.

b.  Reinforcement  is  given  for  each  selected/acceptable  response  only  when  the  push  button  is  being 

pressed. 

c.  The reinforcement counter is incremented for all reinforcements.
d.  The CS lights and tones are available and turn ON at the intensity selected from 1 - 20 only when the 

shaping push button is pressed and held.  CS is turned OFF as soon as the push button is released.  

e.  A single Reset pauses the schedule and turns selected outputs OFF.  The green status light flashes while 

the schedule is paused.  Press Run to continue and Reset to terminate the schedule.  Use the Down button 

to reveal multiple data screens and the Up button to return to previous screens while the schedule is 

running or paused.  Data will be lost as soon as the schedule is terminated.

f.  The Shock is OFF.  

Summary of Contents for 81335C

Page 1: ...Model 81335C 3700 Sagamore Parkway North Lafayette IN 47904 USA Tel 765 423 1505 800 428 7545 Fax 765 423 4111 info lafayetteinstrument com www lafayetteinstrument com 2003 by Lafayette Instrument Company Inc All Rights Reserved Rel 6 16 06 Operant Conditioning Console User s Manual MAN510 ...

Page 2: ...aping activation is provided and student participation encouraged with front panel display of counters latency timers where appropriate and schedule settings The 81335C Behavioral Control Console features a 20 x 2 character LCD display screen membrane touch controls and rear panel connections The complete behavioral control package includes a 30 V DC Power Supply Shaping Push Button 10 Cage Connec...

Page 3: ...nnection cable between the chamber mini I O module and the 81335C Control Connect the shaping push button to the back panel of the control If the shock grid will be used connect its supplied cable to the back panel of the control 2 Auxiliary connections 1 4 may be used for further manipulation Additional instructions for the use of these connections are explained within the auxiliary output contro...

Page 4: ...LIQ SHK AIR Series of tests Scroll up Within Schedule Selection Menu Press once to move up to the test menu from the first item magazine training on the schedule selection list Press again and it will go to the end of the list After scrolling down it will scroll up to see the previous schedule selection Within Schedule Setup Pressed once to change a selection status from N to Y or to increase the ...

Page 5: ... response Count CT Ratio Value in Discrimination 2 4 DRL Interval Value seconds ESCT Escape Count FI Fixed Interval Value seconds FLB Front Left Bar response Count FR Fixed Ratio Value FRB Front Right Bar response Count INT Interval Value s Escape Avoidance INTVL Interval Value seconds in DRH LAT Latency to respond in Escape and Avoidance Schedules LIQ Liquid Reward Count ODL Omnidirectional Lever...

Page 6: ... decrease the timed reward output Use the Shaping push button to activate the reward output as indicated Press Enter to repeat this test for the Liquid Dispenser Shock Output or Air Puff Control Press Reset to return to the Magazine Training menu Magazine Training Step by Step use of the Menu Setup Displays When the control console is first powered up it defaults to the MagazineTraining schedule s...

Page 7: ...pective counters b Reinforcement is given for each selected acceptable response and for each push button activation c The reinforcement counter is incremented for all reinforcements d The CS lights are available and turn ON at the intensity selected from 1 20 when the Run command is issued The CS remains ON for as long as the schedule is running e A single Reset pauses the schedule and turns selec...

Page 8: ...l distribution is quasi gaussian with a lower limit of x mean Although the upper limit is open the probability of obtaining a value greater than 2 x mean is very low Fixed Interval FI a All press bars the Omni directional lever and the shaping push button increment their respective counters b Reinforcement is given for the first selected acceptable response to occur after t seconds have elapsed Th...

Page 9: ...proceed to the next item When Set Up is complete press Run to activate the schedule b Only the Front Right and Front Left Press Bars are used for this schedule Two counters are provided for each to separate responses in S1 vs S2 A counter is also provided for the shaping push button c Reinforcement is given for each selected acceptable response only for each nth response This value is set in the R...

Page 10: ...ifferential Reinforcement of High Response Rates DRH This schedule is just the opposite of the DRL and set up uses menu items from both FR and interval schedules The task is to complete the FR prior to the interval timing out Once the interval times out the FR count is reset to zero a Select acceptable responses and reinforcement options as in previous schedules b Ratio Control Use the Up and Down...

Page 11: ... is terminated g The Shock is OFF Auxiliary Outputs 3 and 4 are available Avoidance AVD Press Enter to select Avoidance from the Schedule Selection list and display the first Set Up screen Use the Up Down buttons to change the setting and or press Enter to accept the default setting and proceed to the next item When Set Up is complete press Run to activate the schedule a One or more acceptable res...

Page 12: ...elected and counters frozen between trials i A single Reset pauses the schedule and turns selected outputs OFF The green status light flashes while the schedule is paused Press Run to continue and Reset to terminate the schedule Use the Down button to reveal multiple data screens and the Up button to return to previous screens while the schedule is running or paused Data will be lost as soon as th...

Page 13: ...nser 5 BK LF STIM LITE Black wire of the Back Left Stimulus Light 6 BK RT STIM LITE Black wire of the Back Right Stimulus Light 7 Not Used 8 FR LF STIM TONE Black wire of Front Left Sonalert Tone Generator 9 FR RT STIM TONE Black wire of Front Right Sonalert Tone Generator 10 Not Used 11 Not Used 12 Air Puff Control Note Other 28 V DC devices may be connected to any programmed output and will be a...

Page 14: ...ided for those wishing to wire directly to a DB 25 Plug connecting to the rear panel socket labeled Chamber Connection Pin Function 1 28 V DC 2 Output FR RT STIM LITE 3 Output LIQUID 4 Output BK RT STIM LITE 5 Output FR LF STIM TONE 6 7 Output AIR 8 9 10 11 Input BK LF PRESS BAR 12 Input FR RT PRESS BAR 13 GROUND 14 Output FR LF STIM LITE 15 Output PELLET 16 Output BK LF STIM LITE 17 18 Output FR ...

Page 15: ...User s Manual Fax 765 423 4111 www lafayetteinstrument com info lafayetteinstrument com 15 Notes Add any user notes here for your convenience ...

Page 16: ...pairs Damaged Goods Damaged instrumentation should not be returned to Lafayette Instrument prior to a thorough inspection If a shipment arrives damaged note damage on delivery bill and have the driver sign it to acknowledge the damage Contact the deliveryservice andtheywillfileaninsuranceclaim Ifdamageisnotdetectedat thetimeofdelivery contactthecarrier shipperandrequestaninspectionwithin 10 days o...