3 LCSU 4 Setup
English
7
Assemble 300 ml Version
1
Push the upper port connection into the Vacuum Inlet and
check that the lower part of the Canister clicks in place.
2
Connect the Patient Tube to the Patient Port. Ensure that
all connections are secure to prevent leakage.
1
2
300ml
250ml
200ml
150ml
100ml
50ml
300ml
250ml
200ml
150ml
100ml
50ml
Important
• The 300 ml Canister is sealed, with an internal Filter. The
Canister is disposable and cannot be cleaned. The Filter
automatically stops suction/flow when the Canister is full,
or the Filter becomes saturated if the unit tips onto its side
during use.
• The 300 ml Canister (Cat. No. 886100) can also be used with
the previous model LCSU 3. However, the LCSU 4 cannot
be used with LCSU 3 Canisters.
• To improve the upright stability, a Wire Stand (Cat. No
886115) is offered as an Optional Accessory. This adds
both a handle and a foot to the unit.
Caution
Do not attempt to install or use LCSU 3 Canisters on
LCSU 4.
Check Before Each Use
1
The Unit should not be damaged.
2
The Unit should be clean.
3
All parts should be properly assembled (Canister, Tubes etc.).
4
Perform Device Test after each reassembly (see Chapter 6).
5
Check Battery level: While performing the Device Test the
Battery level should not illuminate RED. If the Battery
level illuminates RED, the Battery should be charged. See
Charging Instructions (Chapter 5).
Important
For both models - always have an extra Canister available
in case the first Canister is filled completely, or the unit
becomes tipped on to its side and the Filter becomes
saturated and stops the suction/flow.
Warning
If the Canister on either model fills and the shut-off
mechanism activates, and you have no spare Canister
ready for immediate replacement, shut off LCSU 4 and
utilize alternative methods according to local protocol
to clear patient’s airway. Continued efforts to suction
with LCSU 4 Canister(s) filled may cause overflow that
will prevent suctioning, damage the pump, void the unit
warranty, and lead to prolonged downtime.
Cat. No. 886115
Summary of Contents for LCSU 4
Page 2: ......
Page 180: ...180 7 LCSU 4 1 2 DC AC 12V LCSU 4 1 2 3 800 ml 4 300 ml 1 2 2 1 5 2 LCSU 4 Laerdal Medical...
Page 194: ...7 194 LCSU 4 1 2 12V LCSU 4 1 2 3 800 ml 4 300 ml 1 2 1 5 2 LCSU 4 Laerdal Medical...
Page 208: ...7 208 LCSU 4 1 2 12 V DC AC LCSU 4 1 2 3 800 ml 4 300 ml 1 2 1 5 2 LCSU 4 Laerdal...
Page 218: ...4 218 LCSU 4 1 6 12 12 12 12 LCSU 4 12 12 LCSU 4 175 150 200 LCSU 4...
Page 220: ...5 220 886111 LCSU 4 LCSU 4 886112 LCSU 4 P...
Page 222: ...7 222 Laerdal LCSU 4 1 2 12 LCSU 4 1 2 3 800 4 300 1 2 1 5 2 Laerdal Medical...
Page 255: ......
Page 256: ......
Page 257: ......