7 Troubleshooting
English
12
Troubleshooting
Warning
Potential for Electrical shock. Do not attempt to open or disassemble pump or electrical accessories.
LCSU 4 will not power “ON”
1
Verify that the Battery is properly installed and charged.
2
Check alternative 12V DC or AC sources.
LCSU 4 runs, no vacuum, or insufficient vacuum
1
Verify proper Canister and Tubing connections.
2
Check system for leaks in Canister and Tubing connections.
3
Check Canister overflow protection float (800 ml).
4
Check if the Filter has been shut-off (300 ml).
Vacuum too high or too low
Turn Vacuum Regulator to increase or reduce vacuum level.
Battery will not charge
1
Verify that the Battery is installed and connected.
2
Re-connect source power and observe charge (all connections).
Battery seems insufficient
1
Charge the Battery for 5 hours.
2
Perform the Battery Test.
Important
If the LCSU 4 condition is not resolved, contact Laerdal Medical or one of its authorized distributors for advice.
Summary of Contents for LCSU 4
Page 2: ......
Page 180: ...180 7 LCSU 4 1 2 DC AC 12V LCSU 4 1 2 3 800 ml 4 300 ml 1 2 2 1 5 2 LCSU 4 Laerdal Medical...
Page 194: ...7 194 LCSU 4 1 2 12V LCSU 4 1 2 3 800 ml 4 300 ml 1 2 1 5 2 LCSU 4 Laerdal Medical...
Page 208: ...7 208 LCSU 4 1 2 12 V DC AC LCSU 4 1 2 3 800 ml 4 300 ml 1 2 1 5 2 LCSU 4 Laerdal...
Page 218: ...4 218 LCSU 4 1 6 12 12 12 12 LCSU 4 12 12 LCSU 4 175 150 200 LCSU 4...
Page 220: ...5 220 886111 LCSU 4 LCSU 4 886112 LCSU 4 P...
Page 222: ...7 222 Laerdal LCSU 4 1 2 12 LCSU 4 1 2 3 800 4 300 1 2 1 5 2 Laerdal Medical...
Page 255: ......
Page 256: ......
Page 257: ......