3.3.2 Disposable Canister System
The Disposable Canister System consists of:
- A 1000 ml reusable transparent plastic
Canister including a Canister Connector.
- A disposable liner including lid with filter, tube
and connectors.
The Abbott Receptal® Canister contains a
hydrophilic filter that shuts off the vacuum if the
canister is full or the LSU tips over.To restore the
vacuum, replace the inner liner.
Note:
Do not use an Abbott liner without a filter!
The disposable parts must be replaced after each
use.The Reusable Canister must be cleaned as
stated on the canister.The Vacuum Connector
Tubing is reusable and should be cleaned in
accordance with section 6.3.
A Filter Tubing is available for the Abbott Canister
System (see section 8).
3.4 Power Sources
The LSU is delivered with an AC Power Cord for
connection to AC mains and a DC Power Cord
for connection to DC mains.
The device can be operated from the internal
battery, and can be operated or charged from
one of the following external Power Sources:
- AC mains when used with the AC Power
Cord: 100-240 VAC (50/60/400 Hz).
- DC mains when used with the DC Power
Cord: 12-28 VDC.
A Wall Bracket to hold the LSU during operation
and (optional) charging is available separately
(see section 8).
External power connection will cause the
External Power Indicator to be lit.
Always check that it is lit when
connecting external power or
inserting the LSU into the Wall Bracket.
3.5 Automatic Power-save Function
The LSU has an Automatic Power-save Function
that switches the pump motor off.
While in this mode, the Power ON
Indicator will flash slowly
(approx. once per second).
This function will be activated when the
Operating Knob is set to 200, 350 or 500+
mmHg and the actual vacuum level has been
continuously higher than 120 mmHg for more
than 2 minutes.
To exit the Power-save Mode and revert to
normal operation, set the Operating Knob to
any other position and then go back to required
setting.
PATIENT Inlet
Patient Suction Tubing
VACUUM Inlet
Disposable Liner
including lid with
filter, tube and
connectors
Reusable Parts
Reusable Canister
T-bar
Vacuum Connector
Tubing
Disposable Parts
AC Connector Socket
DC Connector Socket