80 mmHg
120 mmHg
200 mmHg
350 mm Hg
500+mmHg
12 sl/min.
16 sl/min.
20 sl/min.
23 sl/min.
> 25 sl/min.
23
ENGLISH
11. Specifications
11.1 Classification
Electrically powered medical suction equipment for field and transport use, according to
ISO10079-1:1999.
High vacuum/high flow.
Not suitable for use in the presence of flammable liquids or gases.
Internally powered/class ll equipment type BF, according to IEC 601-1:1988.
Protection class IP34D*, according to IEC 529:1989.
* - Protected against solid foreign objects of 2,5 mm Ø and greater
- Protected against splashing water
- Protected against access with a wire
11.2 General tolerance
Overall tolerance ±5%
11.3 Physical Characteristics
Size:
315 mm (12,4 in) x 330 mm (13 in) x 160 mm (6,3 in), (h x w x d)
Weight:
4kg (8,9 lbs) (including battery)
Canister Capacity:
1000 ml
Canister graduation
accuracy:
±5% of full scale
Patient Suction
Tubing (non-sterile):
8 mm (0.315 in.) inside diameter x 1.5 m (59 in.) length
11.4 Operation
Approx. free air flow at different settings:
sl = litre at standard atmospheric pressure and temperature.
Approx. battery operation
time (free air flow)
at different settings (±10%):
Approx noise levels (free air flow)
at different settings:
Vacuum - Max.: > 500 mmHg (67 kPa).
Vacuum - Range: 80 - 500+ mmHg (11 - 67 kPa).
Vacuum Indicator accuracy: ±5% of full scale.
80 mmHg
120 mmHg
200 mmHg
350 mmHg
500+mmHg
3h 20 min
2h 20 min
1h 30 min
1h
45 min
80 mmHg
120 mmHg
200 mmHg
350 mmHg
500+mmHg
46 dBA
48 dBA
51 dBA
53 dBA
56 dBA