6.4.5 Disinfection

To disinfect the parts, the following methods are recommended:

6.5 Charge the Battery

The internal rechargeable Lead-Acid battery can

be charged directly from external AC or DC

power as described in section 3.4. No external

charger is needed.To charge the battery, proceed

as follows:

1. Make sure the Operating Knob is set to "0".

Connect either external AC or DC power to 

the LSU and charging will start automatically.

2. During charging the Battery Status Indicator 

will indicate approx. achieved battery capacity 

(see section 3.2.2.5 for details).

3. The LSU must be placed on charge for a 

minimum of 24 hours to reach full battery 

capacity.The fast charging gives approx.

80% capacity after 3 hours 

(for a new battery).

Repetitive charging for only 3 hours is 

not recommended.

Note:

- To maintain satisfactory operation of the battery,

it is recommended to place the LSU on 

continuous charge immediately after use.

- It is recommended to always fully charge the 

battery. Repetitive charging to a lower capacity 

level will reduce battery life.

- If it is not possible to place the LSU on 

continuous charge when not in use, make sure 

the battery is charged for a minimum of 

24 hours at least once a month.

-  Do not store the battery when it is discharged.

Always fully charge the battery before storage.

- The recommended ambient temperature for 

charging is from 15°C to 25°C.

- The battery will not be charged when the LSU 

is in operation.

- If no battery is installed, the battery status 

indicator will be fully lit for approx. 5 sec. until 

it is turned off.

6.5.1 External Battery Charge

The LSU battery can also be charged 

in an optional External Battery Charger.

See section 8.

6.6 Check the Battery Quality

If poor battery quality is 

suspected, charge the battery 

for a minimum of 24 hours and 

then perform the following test 

without connection to external 

power:

- Run the Device Test and then let the LSU 

operate continuously at 500+ mmHg setting 

and free air flow for 20 minutes.

- If the LSU stops before completing the 20

minutes, the battery should be discarded.

Method

Glutaraldehyde

Sodium Hypochlorite,
(not cleared 
for use in the US)

Virkon

Steam autoclaving

Temperature/
Concentration

Room temperature/ 
Conc.: 2%

Room temperature/ 
Conc.: 0.5% 

Room temperature/ 
Conc.: 1%

Autoclave at  
max. 121° C

Contact time 

60 minutes

20 minutes

10 minutes

20 minutes

Post-treatment

Rinse all parts in clean,
warm water. Allow to dry.

Rinse all parts in clean,
warm water. Allow to dry.

Rinse all parts in clean,
warm water. Allow to dry.

Allow parts to cool.

LED 4____

LED 3____

LED 2____

LED 1____

Summary of Contents for 78 00 00

Page 1: ...Directions for Use Laerdal Suction Unit www laerdal com 78 00 10 78 00 00...

Page 2: ...le to place the LSU on continuous charge when not in use make sure the battery is charged for a minimum of 24 hours at least once a month Do not store the battery when it is discharged always fully ch...

Page 3: ...lace the Disposable Canister 13 5 3 Assembly of the Disposable Canister System 14 5 4 After Use 15 6 Maintenance 15 6 1 General 15 6 2 Battery 15 6 3 Cleaning the Cabinet 15 6 4 Decontamination of Reu...

Page 4: ...m levels are usually selected for tracheal suctioning and the suctioning of children and infants 2 3 Contraindications None known 2 4 Features and Benefits Splashproof rugged and solid Easy access to...

Page 5: ...nt Suction Tubing Suction Catheter Adapter Holder for Water Bottle Patient Suction Tubing Water Bottle Canister Lid with Float Ball Battery Door Release Arm Canister 3 1 2 LSU w Disposable Canister Sy...

Page 6: ...it continuously while the LSU is switched ON It flashes rapidly approx twice per second during the Device Test It flashes slowly approx once per second while the Automatic Power save Function is activ...

Page 7: ...ately after switching from external power to internal battery operation all 4 LED s will flash for 5 seconds before the remaining battery capacity is displayed CAPACITY for a new battery stored in roo...

Page 8: ...d without the Aerosol Filter or the Float Ball It is recommended always to have extra Aerosol Filters with the LSU in case one has to be discarded The Aersol Filter is not intended for microbio logica...

Page 9: ...ng external Power Sources AC mains when used with the AC Power Cord 100 240 VAC 50 60 400 Hz DC mains when used with the DC Power Cord 12 28 VDC A Wall Bracket to hold the LSU during operation and opt...

Page 10: ...ting Knob to the required vacuum level The LSU will automatically be switched ON and start to operate The green Power ON Indicator is lit continuously while the LSU is switched ON 6 When suction is co...

Page 11: ...yellow Gasket in the Lid make sure that the gasket is placed flat in the Lid and not twisted 2 If the Float Ball has been removed hold the lid upside down and snap the ball into place as shown Note Ne...

Page 12: ...olor code Blue on Blue Yellow onYellow 4 3 1 Reusable Canister System Quick reference The assembly of the tubing is illustrated on the right side of the LSU 1 Connect the blue Aerosol filter to the bl...

Page 13: ...erated in the upright position to prevent overflow of suctioned material 7 Set the Operating Knob to the required vacuum level and the LSU will automatically be switched ON and start to operate The gr...

Page 14: ...to the blue Vacuum Inlet on the LSU Color Code Blue on Blue 4 Insert the Canister in the in the Canister Holder and slide it into the LSU 5 Connect the yellow Connector on the Vacuum tubing to the Ye...

Page 15: ...ed both on a monthly basis and after each use 6 2 Battery To maintain satisfactory operation of the battery it is recommended to place the LSU on continuous charge when not in use Replace the battery...

Page 16: ...eparate parts as indicated in the picture above The Float Ball on the reusable canister can be snapped out of the lid 6 4 3 Rinsing Rinse all parts under cold running water Immerse in warm water 30 40...

Page 17: ...24 hours at least once a month Do not store the battery when it is discharged Always fully charge the battery before storage The recommended ambient temperature for charging is from 15 C to 25 C The...

Page 18: ...is user initiated test program for all models which can identify whether the LSU operates satisfactorily or if the LSU needs service The program runs 4 different tests 1 It tests for occlusions in th...

Page 19: ...est results turn the Operating Knob to 0 to exit the Device Test Note If you need to interrupt the test and revert to nor mal operation turn the Operating Knob to another position and then select the...

Page 20: ...LED 2 lit up LED 3 lit up LED 4 lit up Test result indications Test Passed 100 mmHg Test Passed 300 mmHg Test Passed 500 mmHg Test Passed 450 mmHg Action if test failed Check possible blockages e g tw...

Page 21: ...ter adapter pkg 10 78 40 00 LSU Reusable Canister 78 04 30 LSU Reusable Canister Holder 78 10 06 Angled Connectors pkg 10 78 10 02 Float Ball pkg 10 78 40 07 Vacuum Seal 78 40 08 Gasket 79 35 00 Water...

Page 22: ...n the Abbott canister is leaking air into the tubes The Abbott liner is leaking Battery is not charged Tube s is kinked or twisted The external Power indicator is not lit ACTION Check power cord conne...

Page 23: ...e Overall tolerance 5 11 3 Physical Characteristics Size 315 mm 12 4 in x 330 mm 13 in x 160 mm 6 3 in h x w x d Weight 4kg 8 9 lbs including battery Canister Capacity 1000 ml Canister graduation accu...

Page 24: ...front Styrene Ethylene Betyl Styrene SEBS Cabinet back PC ABS Cabinet base w Protector PC ABS SEBS Battery door SEBS Connector retainer for battery Poly Oxy Methylene POM Operating Knob POM Rotor for...

Page 25: ...rovided in this section Portable and mobile RF communication equipment can affect MEDICAL ELECTRICAL EQUIPMENT Warning MEDICAL ELECTRICAL EQUIPMENT should not be used adjacent to or stacked with other...

Page 26: ...30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Mains power quality...

Page 27: ...cy rangeb Interference may occur in the vicinity of equipment marked with the following symbol Guidance and manufacturer s declaration electromagnetic immunity NOTE 1 At 80 MHz and 800 MHz the higher...

Page 28: ...2005 Laerdal Medical AS All rights reserved 8179 Rev A Printed in Norway 0434...

Reviews: