Seite / Page:  10

 

When the equipment is set up care must be taken 

that  it is standing  horizontally  and  is  well  seated 

on all four bearing pads.

 

 

For security reasons, screen clamp and the all

-

in

-

one  PC  dismantled and  shipped separately.  The 

screen mounting screw on the test bench and fit 

the all

-

in

-

one PC on the holder.

 

 

 

Connect the mains cable of the apparatus with a 

220

 

V / 50

--

60

 

Hz mains outlet. Turn on the appa-

ratus.

 

For  testing  in  the  MP  range  the  MP  line  of  the 

breathing apparatus in the Euro connection is dis-

connected  and  connected  with  the  MP  connec-

tion points 

(H) 

and 

(I).

 If the breathing apparatus 

to be tested is not supplied with a Euro connector 

it is necessary to use an appropriate test adapter. 

Please approach the manufacturer of the breath-

ing apparatus to find which adapter you need.

 

 

 

 

Installation of the Labtronic NT Software

 

The operating system and software Windows 

XP  pro  (Windows7  from  10.2010)  and  the 

software  Labtronic.NT  (Labtronic.NET  from 

10.2010  )  are  preinstalled.  By  switching  on 

the test, the computer is booted and Labtron-

ic.NET initialized.

 

After  entering  the  username  and  the  pass-

word, you can begin the testing.

 

 

Beim  Aufstellen  darauf  achten,  dass  das  Gerät 

waagrecht  steht  und  an  allen  4  Auflagepuffern 

gut aufliegt. 

 

 

Aus  Sicherheitsgründen  werden  Bildschirmhalte-

rung  und  der  all

-

in

-

one  PC  demontiert  und  ge-

trennt  geliefert.  Die  Bildschirmhalterung  an  das 

Prüfgerät anschrauben und den all

-

in

-

one PC auf 

die Halterung montieren.

 

 

Geräteanschlusskabel mit einer 220 V /50

-

60 Hz 

Stromquelle verbinden. Gerät einschalten. 

 

 

Für  Prüfungen  im  Mitteldruckbereich  wird  die 

Mitteldruckleitung  des  Pressluftatmers  an  der 

Euro

-

Kupplung  abgekuppelt  und  mit  den 

Mitteldruckanschlüssen 

(H) 

und 

(I) 

verbunden. 

Besitzt  der  zu  prüfende  Pressluftatmer  keine 

Euro

-

Kupplung,  so  muss  ein  entsprechender 

Prüfadapter verwendet werden. Bitte wenden Sie 

sich an den Gerätehersteller des Pressluftatmers.

 

 

 

 

Installation der Software Labtronic NT

 

Das  Betriebssystem  Windows  XP  pro  (ab 

10.2010 Windows7 pro) und die Software Labtro-

nic.NT  (Labtronic.NET  ab  10.2010)    sind  bereits 

vorinstalliert.  Mit  dem  Einschalten  des  Prüfstan-

des  wird  der  Computer  gebootet  und  Labtro-

nic.NET initialisiert.

 

Nach  Eingabe  des  Benutzernamens  und  des 

Kennwortes kann mit dem Prüfen begonnen wer-

den.

 

9.  Inbetriebnahme

 

9.  Starting Up

 

Summary of Contents for Labmatic 400.COM

Page 1: ...ätebeschreibung Description of the equipment 2 3 Sicherheitshinweise Safety information s 3 4 Technische Daten Specifications 4 5 Bedienelemente Operation controls 5 6 Geräteanschluss Device connection 6 7 Bauteilebeschreibung Explanation of the constr units 7 8 Zubehör Accessories 8 9 9 Inbetriebnahme Starting up 10 10 Wartung und Pflege Care and Maintenance 11 Anhang zur Betriebsanleitung für Pr...

Page 2: ...mitted by Labtec Otherwise the sale and delivery conditions of Lab tec GmbH apply which are in no wise extended by these notes 2 Gerätebeschreibung Labmatic 400 COM ist ein Prüfgeräte zum Prüfen von Atemschutzmasken nach EN 136 Pressluftatmern nach EN 137 und gasdichten Chemikalienschutz anzügen CSA Die erforderliche Luftversorgung erfolgt über eine 300 oder 200 bar Atemluftflasche mittels speziel...

Page 3: ...15 A fuse cut out is integrated in the connec tion power socket on the back of the machine This protects the test stand against extreme incom ing power surges There is a spare fuse in the power socket Intrinsic protection The test stand has extensive intrinsic protection against operating error If the pressure in a partic ular pressure sector rises beyond the limit value entered for example the sy...

Page 4: ... x VGA Software Operating system Windows XP pro Service Pack 3 from 10 2010 Win dows 7 Test and operation software Lab tronic NT complete installed from 10 2010 Labtronic NET HD Druckluftan schluss Rohrverschraubung 6 mm S für zentrale 300 bar Luftversorgung und HD Kupplung für wahlweise 300 oder 200 bar FL Zuleitung HD Sensor Messbereich Signal Nulloffset Linearität Hysterese 0 400 bar 4 20 mA 1 ...

Page 5: ...mit Messzelle D Hochdruckanschluss F Geräteanschluss 300 bar G CSA Füllanschluss H Mitteldruck Lungenautomatenanschluss I Mitteldruck Geräteanschluss J Netzanschluss mit Schalter A Housing B Touch screen computer C Test head with measuring point D High pressure connection F HP connection for appliances G Connection for inflation of CPS H MP connection for demand valves I MP connection for applianc...

Page 6: ...ion 4 LED 24 V present 5 LED 5 V present 6 Power connection with switch 6 Montage des Touch Screen PC Mounting the Touch Screen PC 1 TFT LCD Wandhalterung an Prüfstandgehäuse anschrauben Schrauben mit Zierkappen abdecken 2 TFT LCD Wandhalterung an LED Touch Screen PC anschrauben 3 Trafo mit Klettband am Gehäuse befestigen 4 Kaltgeräteverlängerung und USB Kabel mit Gewebeschlauch befestigen 5 Kabel...

Page 7: ...tiert Das andere Endstück enthält einen Nippel für die am Prüfstandgehäuse angebrachte Hochdruckkupplung Achtung Niemals Leitung unter Druck ein oder auskuppeln CSA Prüfanschluss F Schnellkupplung zum Befüllen von CSA Schnittstelle An der linken Gehäuseseite befindet sich eine USB Schnittstelle für die Datenübertragung an einen PC Test bench The ergonomic design means that the test station can als...

Page 8: ...IN 58 600 136 020 Stoppuhr für Dichtigkeitsprüfungen 808 605 Silicon Spray Zur Pflege des Prüfkopfes 142 502 Prüfkopf AC 500 Ersatzprüfkopf im Austausch Weitere Adapter für fabrikatsabhängige Steckanschlüsse auf Anfrage Order No Description Explanation 161 027 Dust cap RD 40 Dust cap for full face masks with screw 40 x 1 7 161 028 Dust cap PE 45 Dust cap for full face masks with screw M 45 x 1 5 1...

Page 9: ...85 CSA Adapter KP CSA mit Kappler Überdruckventilen 161 086 CSA Adapter RP CSA mit Respirex Überdruckventilen 8 Zubehör 8 Accessories Order No Description Explanation 161 061 CPS Test Adapt er Type CGA 1500 For testing of chemical protection suits basic unit in connection with CPS adapters 161 062 CPS Adapter AU CPS with MSA over pressure valves 161 063 CPS Adapter DG CPS with Dräger over pressure...

Page 10: ...g on the test the computer is booted and Labtron ic NET initialized After entering the username and the pass word you can begin the testing Beim Aufstellen darauf achten dass das Gerät waagrecht steht und an allen 4 Auflagepuffern gut aufliegt Aus Sicherheitsgründen werden Bildschirmhalte rung und der all in one PC demontiert und ge trennt geliefert Die Bildschirmhalterung an das Prüfgerät anschra...

Page 11: ...t oder aus anderen Gründen nicht mehr dicht ist so kann der Prüfkopf im Austauschverfahren gewechselt werden Bitte senden Sie den Prüfkopf ins Werk Sie erhalten umgehend einen Austauschkopf The test stand does not need any special care It should however be protected from external harm ful influences water damp in the form of steam dust The test head needs no special maintenance but should be regul...

Page 12: ...by certify that the above mentioned product meets the requirements set forth in EEC Guideline 89 392 and EMC 89 336 Class A including all changes and addendums to date thereto The above mentioned product meets the following stand ards and guidelines Nous Vous Confirmons que l appareil cité ci dessus correspond aux exigences des directives 89 392 CEE ainsi qu à la directive CEM 89 366CEE Classe A c...

Reviews: