35
5.
On Channel B’s Logic tab, select: inB = outA or not in1. This routes the
button push signal and the delayed signal to channel B.
6.
Since you’ve made changes to multiple channels, click the “Program All
Settings” button.
Now the test is ready. Depress and release the trigger button quickly. Channel A
will produce a trigger pulse 500 ms later. Both the button push and delayed
signal will appear on the inputs of Channel B. When Channel B receives the
button-push signal, it will begin its count. When Channel B receives the delayed
pulse from Channel A, it will calculate the time difference (500 ms), multiply it
by 32/16, and subtract the 200 µs offset. 999.8 ms after receiving the signal from
Channel A, Channel B will send a 100 ms output pulse.