29
Inverted Output
Check the box to set the output high until a valid
trigger.
On the
Triggering
tab choose:
Normal
The channel can retrigger an unlimited number of
times
Skip N triggers
The channel triggers only after the specified number
of valid input pulses (between 0 and 2,147,483,647).
Retrigger N times
Limits the number of retriggers allowed between
resets (between 0 and 2,147,483,647). To allow a
channel to be triggered again after reaching the limit,
the channel must be reset. To reset both channels,
press the “Reset” button on the LC220 or choose
Actions>Reset in Trig220. To reset only one
channel, click the “Reset” button next to the channel
settings.
Trigger on input high
While the input is high, channel can be repeatedly
triggered.
Trigger on rising input
Sets the channel to be edge triggered (i.e., an input
signal that stays high will only trigger the channel
once)
.
The panel LED
for the noise suppression channel will glow green until an input
signal is detected. It will then glow red until the duration period has elapsed. At
this point, the LED will respond as follows:
Retriggering
enabled?
Edge
trigger?
LED
no
yes or no
remains red (no more triggers accepted)
yes
no
glows green (waiting for next trigger)
yes
yes
is off; glows green when trigger goes low