2.5 Setting Up Consumables and Preparing the
Patient
Synergy uses
Consumable Traceability
to track consumables used during studies. Prior
to beginning consumable setup, ensure to start a study. For instructions on starting the
study refer to the
Start Study Button
section on page
56
. For more information on
Consumable Traceability
refer to the
Procedure – Preview Phase
section on page
57
.
Complete consumable and patient setup in the preview phase of the study. Consumable
information will show in the
Consumable Traceability
tab available in the
Preview
window once entered or scanned (Figure 6).
Figure 6: Consumable Traceability – Preview Window
Infusion Pump Tubing Setup
LABORIE Infusion Pump Tubing is intended for single use only. Do not re-use, reprocess, or
resterilize pump tubing. Use of the Infusion Pump Tubing outside of its intended use may
affect pump accuracy. Do not run the pump at a speed exceeding the limitation of the
catheter.
IMPORTANT
:
•
Always use only LABORIE Infusion Pump Tubing to ensure a properly and
accurately functioning system.
•
Catheters, the saline bag, and the Infusion Pump Tubing are consumables
intended for single use with a single patient only; they must be replaced after
every use.
•
To maintain patient safety during normal use, the maximum pump speed is
preset to 150ml/min.
•
Always refer to the Instructions for Use provided with the consumable deployed
for information on proper use of the device.