background image

 
BTA 060, Rev 1K1 

Page 10/18 

6.2  Main menu 

The main menu contains the following functions: 

Display 

Designation 

Description 

 

Basic functions 

Setting the unit, setting the zero point, 
min./max. pointer 

 

Display functions 

All settings relating to the display 

 

Switch point settings 

Configuration of the switching outputs 
(optional) 

 

System data 

Displaying system data (versions, serial 
number); reset to factory settings 

 

Return 

Return to display mode 

Table 3:  The items in the main menu 

6.3  Basic menu (

bASE

The basic menu contains the following items: 

Display 

Designation 

Description 

 

Measuring unit 

Setting the measuring unit via a parameter 
list 

 

Set zero point  
(Teach lower range) 

Setting the applied pressure as zero point 
(0 bar) 

 

Min. pointer (low) 

Display resp. delete the min. pointer 

 

Max. pointer (high) 

Display resp. delete the max. pointer 

 

Return 

Return to the main menu 

Table 4:  The items in the basic menu 

Summary of Contents for PASCAL CS

Page 1: ...Table of contents 1 General Information 2 1 1 Intended Use 2 1 2 General Safety Notes 2 1 3 CE Marking 2 2 Transportation and Storage 2 3 Installation and Commissioning 2 3 1 Electrical Connection 3 3...

Page 2: ...ble EU Directives for placing products on the market within the European Union The following guideline applies to model series PASCAL CS21xx EMC Guideline EMC 2004 108 EG 2 Transportation and Storage...

Page 3: ...switching output 2 Pgm1 2 programming pins Figure 1 Pin assignment for M12 circular connector device side 3 1 1 Connecting the switching outputs optional The switching outputs are potential free They...

Page 4: ...nless steel housing with one hand and turn the display unit with the other hand into the wanted position The turning angle is limited by an internal limit stop Do not try to force the display unit bey...

Page 5: ...configure it with a communication modem and the PC Software COMLINE S This is described in a separate manual 4 2 Maintenance When properly installed in accordance with applicable specifications this...

Page 6: ...y the LED flashing When there is no button pressed the output states are displayed 6 1 2 Display mode Measured value screen When the device is switched on it goes into display mode The currently measu...

Page 7: ...to display mode At the end of a menu typically the rEt item you return to the first menu item by repressing the down arrow button again Similarly you can jump from the first menu item to the end of th...

Page 8: ...the rEt item which allows you to return to the next higher menu level The middle button stores your selection Stor appears on the display to confirm that the value has been stored and the device retur...

Page 9: ...the previous value You confirm the selected digit with the middle button and proceed to the next digit You can cancel the value entry at any time by pressing both arrow buttons simultaneously The dev...

Page 10: ...system data versions serial number reset to factory settings Return Return to display mode Table 3 The items in the main menu 6 3 Basic menu bASE The basic menu contains the following items Display D...

Page 11: ...of the measuring range with the teach lower range menu item tlr When the menu item is selected the entry points to rEt To trigger the function go to YES with one of the arrow buttons and confirm with...

Page 12: ...m in this menu Resetting a pointer is confirmed by showing on the display Display Designation Description Value of min max pointer Value of min max pointer in the selected measuring unit Clear Reset t...

Page 13: ...two ways either by pressing the middle button in display mode or alternately with the measured value When displayed alternately with the measured value the display periods for the measured value and...

Page 14: ...shown on the display will appear flashing e g 99 99 when two decimal places are set for a measured value of 110 mbar 6 5 Switch point menu The switch point menu contains the functions for setting the...

Page 15: ...hen you select a frame function the menu items for switch point and reset point are replaced by the upper and lower frame limits The minimum difference of the frame limits is also 0 5 of the measuring...

Page 16: ...If the pressure is above the switch point the switch is open At the lower range limit the switch is closed Frame normally open Inside of the frame the switch is closed At the lower range limit the swi...

Page 17: ...enu Due to the limited number of alphanumerical segments on the display hardware and software versions are split into two separate items and serial number into four items 6 6 2 Reset to factory settin...

Page 18: ...Display direction 0 normal 180 turned Setting the decimal places zero to three fixed decimal places or automatic Configuring the switching outputs only if available Switch point or upper frame limit o...