Technical specifications 

Description

Specification

Temperature range

-90 °C to 

65 °C ± 2,5 °C

Storage volume

87 L

DxWxH (internal)

510 x 440 x 390 mm

DxWxH (External)

716 x 650 x 862,5 mm

Shelf Dimensions

475 x 440 mm

Required power supply

one phase, 230 V, 50 Hz + earth

Power consumption max./norm.

980/400 W

Weight

106 kg

Refrigerant

R507 / R170

Noise level

45 dB(A)

11 

Safety and supervisory accessories 

For increased safety & supervision of the freezer it can be equipped with the optional accessories 
mentioned below:

CO

back up

Ensures low freezer temperature is maintained after freezer breakdown.

The  CO

2

back  up  has  its  own  temperature  sensor,  battery,  and  power  supply, 

and is fully independent of the freezer’s control system.

Remote alarm

Permits supervision of the freezer by remote CTS.

Printer

Records various  status  messages  of  operational  and  alarm  conditions 
from the control system of the freezer.

12 

External alarm installation 

The external alarm is connected to the socket on the rear of the freezer. 

The external alarm signal is provided by a potential-free switchover contact in the controller. 

It  is  activated  by  the  following  alarm  conditions  Power  failure,  High  Temperature,  and  Low 
temperature. 

13 

Accessories for Snowbird freezer 

8.005.121.011

CO2-Back up

8.005.121.012

Printer

8.005.121.013

RS 485 Interface

8.005.121.014

Extra shelf

8.005.121.015

2/3 Rack for 12 x 2”/3” boxes

8.005.121.016

Full Rack for 18 x 2”/3” boxes

8.005.121.017

Paper for 7 days recorder ( package of 100)

8.005.121.018

7-day recorder

Summary of Contents for ScanCool Snowbird

Page 1: ...Revision 01 Date 2012 11 15 Cat No 9 900 900 752 INSTRUCTION MANUAL Snowbird PERSONAL ULTRA FREEZER...

Page 2: ...ls used in this manual Note Used to direct attention to a special item Warning Used in case of danger of a serious accident or lethal injury LaboGene ApS Industrivej 6 8 Vassinger d DK 3540 Lynge Denm...

Page 3: ...4 Description 5 5 Start up 5 Switch power on 5 5 1 Start up the freezer 6 5 2 Cooling commences 6 5 3 6 Operation 6 7 Programming and default settings 6 8 Maintenance 7 9 Trouble shooting 8 10 Techni...

Page 4: ...3 Installation Note The operating temperature of the freezer compartment is pre set to 80 C at the factory which is the optimum operating temperature with regards to the environmental conditions sound...

Page 5: ...5 1 Insert the battery into battery holder See section 4 Connect the main cable to freezer Note The battery will be charged Front panel User interface and display Air grille with filter Door handle w...

Page 6: ...play will indicate the actual temperature Cooling commences 5 3 Ensure that the door is closed It will take approximately 2 5 hours to lower the temperature of the freezer to the pre set operating tem...

Page 7: ...of the set point 1 Press Enter 2 Press Up or Down till code 1 is displayed 3 Press Down until 87 is displayed 4 Press Enter to store the new value 8 Maintenance Dust filter Vacuum cleaning in very dus...

Page 8: ...from the freezer switch off the power allow all the ice to melt wipe dry and switch on the power again How to check CO2 weight Weigh the CO2 flask and compare it with the Tare weight of the flask A f...

Page 9: ...s its own temperature sensor battery and power supply and is fully independent of the freezer s control system Remote alarm Permits supervision of the freezer by remote CTS Printer Records various sta...

Page 10: ...nt for measurement control and laboratory use EMC requirement General requirements EN 60335 2 24 Safety of household and similar electrical appliances Particular requirements for refrigerating applian...

Reviews: