 9.0  Operating  Instructions

9.1 Constant Operation Protocol

The Constant Voltage, Current (250V and 300V), or Power Mode (250V Only) allow you to specify a 
voltage limit, current limit, and power limit to be used during the entire duration of electrophoresis. 
Review the guidelines provided in this manual before starting electrophoresis using the Labnet Enduro™ 
Power Supplies. 

A basic Constant Voltage, Constant Current, or Constant Power operating procedure of the Labnet 
Enduro Power Supplies is provided below. We recommend reading the guidelines provided in this 
manual for best results before starting an operation. The figures in section 9.1 are examples from 
the 300V display screen. Please note that the 250V model has the same functionality herein with the 
addition of constant power as viewed within the NOTE section of 9.1.

1.  Press the Power Switch on the rear side of the instrument to turn on the Labnet Enduro Power 

Supply. Upon start-up, the Display Screen on the front of the instrument will illuminate. The factory 
default settings (or last settings used) will be displayed.

300V

TIME:00:00

AMP:400mA

 



The chosen constant parameter (Voltage or Current) is displayed in bold on the left side of the 
display.

 



The Timer is the first line on the top right, and the non-constant value is displayed in the second 
line on the right side of the display screen.

2.  Press Constant button to choose constant volt or amperage. 

 



Choosing constant VOLT, the VOLT setting will be large. Amperage defaults to “maximum setting” 
and TIME will be “0”.

300V

TIME:00:00

AMP:400mA

 



Choosing constant AMP, the AMP setting will be large. VOLT will be “maximum value” and TIME 
will be “0”.

400mA

TIME:00:00

VOLT:300V

3.  Press Mode to choose the large VOLT/AMP value on the left side of the display. VOLT/AMP is blinking, 

press the Up or Down Arrow to increase or decrease value. 

NOTE:

 Parameters will blink for 5 seconds. 

After 5 seconds, if you don’t press any button, the parameters will stop blinking and the value will be 
entered.

400mA

TIME:00:00

VOLT:300V

4.  Press Mode. TIME is blinking, press the Up or Down Arrow to increase or decrease value. Time set to 

00:00 means “continuous”, and the timer will count up. Time can be set for 00:01 to 99:99. The timer 
will count down until the time is up and then stop the power supply.

400mA

TIME:00:00

VOLT:300V

5.  Press Mode. Smaller AMP/VOLT on the right side of the display is blinking, press the Up or Down 

Arrow to increase or decrease the value.

400mA

TIME:00:00

VOLT:300V

6.  Press Start/Pause to start the power supply. The parameters will stop blinking and the “run” LED will 

illuminate. “CONSTANT VOLT/AMP” will show on the display.

400mA

TIME:00:00

VOLT:300V

CONSTANT AMP

Summary of Contents for Enduro E0203

Page 1: ...et Enduro Power Supplies Instruction Manual This manual is available in additional languages at www labnetlink com Catalog Numbers High Amperage 250V E0203 E0203 230V General Purpose 300V E0303 E0303...

Page 2: ...ion of Buttons and Switches 3 7 0 Getting Started 4 8 0 Recommendations 4 9 0 Operating Instructions 5 9 1 Constant Operation Protocol 5 9 2 Basic Programming Protocol 250V only 6 10 0 Choosing Limiti...

Page 3: ...nt and power range of the Power Supply To prevent accidental electric shocks to the user always use gel box systems that have safety lids Labnet recommends against the use of gel box systems and or po...

Page 4: ...300W 90W Output voltage range 5 to 250V 2 to 300V Output current range 10 to 3000 mA 4 to 500 mA Operating modes Constant voltage Constant current Constant power 1V step 10 mA step 1W step 1V step 1...

Page 5: ...t Enduro 300V Power Supply is designed to efficiently handle multiple electrophoresis gel tanks and use a small amount of lab space This instrument is ideal for DNA RNA electrophoresis and mini PAGE g...

Page 6: ...the power switch and the AC inlet located on the rear of the unit are easily accessible 4 Ensure the AC power switch is in the Off position 5 Attach the power cord to the AC inlet Use only properly gr...

Page 7: ...e first line on the top right and the non constant value is displayed in the second line on the right side of the display screen 2 Press Constant button to choose constant volt or amperage Choosing co...

Page 8: ...ided below We recommend reading the guidelines provided in this manual for best results before starting an operation The Programmable Mode allows you to vary levels in voltage V current mA and power W...

Page 9: ...ethods resistance increases throughout the run Limiting the voltage provides the following advantages Current and power decrease throughout the run providing an improving margin of safety over time Th...

Page 10: ...t the condition by making sure the tape is removed from the pre cast gel or buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit High voltage applic...

Page 11: ...ent P I x V 13 0 Common Errors Found with Electrophoresis Power Supplies No load The electrophoresis system is not connected to the power leads check the power leads The electrophoresis system has a s...

Page 12: ...orning of any such defect Corning is not liable for any incidental or consequential damages commercial loss or any other damages from the use of this product This warranty is valid only if the product...

Page 13: ...___________________________________ 16 0 Equipment Disposal According to Directive 2012 19 EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment...

Page 14: ...mation visit www corning com lifesciences or call 800 492 1110 Outside the United States call 1 978 442 2200 or contact your local Corning sales office Corning Incorporated Life Sciences 836 North St...

Reviews: