|        LABNET ENDURO POWER SUPPLIES

 11.0  Troubleshooting 

Review the information in the table below to troubleshoot operating problems.

Problem

Possible Cause

Solution

LCD screen remains blank, 
and the fan does not run 
when the power is turned 
on

AC power cord is not 
connected.

Check AC power cord connections at both ends. 
Use the correct cords.

The fuse has blown.

Replace the fuse.

Operation stops with 
alarm: The screen displays 

NO LOAD

Electrophoresis leads 
are not connected to the 
power supply or to the 
electrophoresis unit(s).

  Check the connections to the power supply 

and your electrophoresis cell to make sure the 
connection is intact. 

  Check the condition of the wires in your 

electrophoresis unit. 

  Close the circuit by reconnecting the cables. 
  Press 

START/PAUSE

 to restart the run.

A broken circuit in the 
electrophoresis cell.

Operation stops with 
alarm: The screen displays 

NO LOAD

High resistance due to 
tape left on a pre-cast 
gel, incorrect buffer 
concentration, or 
incorrect buffer volumes 
in the electrophoresis 
cell.

Correct the condition by making sure the tape 
is removed from the pre-cast gel or buffers 
are prepared correctly, and the recommended 
volume of buffer is added to the electrophoresis 
unit.

High voltage application 
is set to run on a very 
low current.

Disable “NO LOAD” alarm on the Display Screen. 

Operation stops with 
alarm: Display shows 

OVER VOLTAGE

Circuit is interrupted.

  Verify the running buffer is correct.
  Verify all cables are attached correctly.
   Turn the Power switch off and on again; 

restart the application.

  If you cannot restart the instrument, turn off 

the power, disconnect the power cord from 
the outlet, and contact Scientific Support.

Operation stops with 
alarm: Display shows 

LEAKAGE

Ground leak detected 
during the run.

  Check the electrophoresis system for improper 

grounding. 

  Restart the power supply by turning the Power 

switch off and on.

Operation stops with 
alarm: Display shows 

OVER TEMP

Power supply is 
overheating.

  Turn off the power supply. Check for sufficient 

airflow around the power supply fan. After 
cooling down, restart the power supply by 
turning the Power switch to the On position. 

  If you cannot restart the instrument, turn off 

the power, disconnect the power cord from 
the outlet, and contact Scientific Support.

Summary of Contents for Enduro E0203

Page 1: ...et Enduro Power Supplies Instruction Manual This manual is available in additional languages at www labnetlink com Catalog Numbers High Amperage 250V E0203 E0203 230V General Purpose 300V E0303 E0303...

Page 2: ...ion of Buttons and Switches 3 7 0 Getting Started 4 8 0 Recommendations 4 9 0 Operating Instructions 5 9 1 Constant Operation Protocol 5 9 2 Basic Programming Protocol 250V only 6 10 0 Choosing Limiti...

Page 3: ...nt and power range of the Power Supply To prevent accidental electric shocks to the user always use gel box systems that have safety lids Labnet recommends against the use of gel box systems and or po...

Page 4: ...300W 90W Output voltage range 5 to 250V 2 to 300V Output current range 10 to 3000 mA 4 to 500 mA Operating modes Constant voltage Constant current Constant power 1V step 10 mA step 1W step 1V step 1...

Page 5: ...t Enduro 300V Power Supply is designed to efficiently handle multiple electrophoresis gel tanks and use a small amount of lab space This instrument is ideal for DNA RNA electrophoresis and mini PAGE g...

Page 6: ...the power switch and the AC inlet located on the rear of the unit are easily accessible 4 Ensure the AC power switch is in the Off position 5 Attach the power cord to the AC inlet Use only properly gr...

Page 7: ...e first line on the top right and the non constant value is displayed in the second line on the right side of the display screen 2 Press Constant button to choose constant volt or amperage Choosing co...

Page 8: ...ided below We recommend reading the guidelines provided in this manual for best results before starting an operation The Programmable Mode allows you to vary levels in voltage V current mA and power W...

Page 9: ...ethods resistance increases throughout the run Limiting the voltage provides the following advantages Current and power decrease throughout the run providing an improving margin of safety over time Th...

Page 10: ...t the condition by making sure the tape is removed from the pre cast gel or buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit High voltage applic...

Page 11: ...ent P I x V 13 0 Common Errors Found with Electrophoresis Power Supplies No load The electrophoresis system is not connected to the power leads check the power leads The electrophoresis system has a s...

Page 12: ...orning of any such defect Corning is not liable for any incidental or consequential damages commercial loss or any other damages from the use of this product This warranty is valid only if the product...

Page 13: ...___________________________________ 16 0 Equipment Disposal According to Directive 2012 19 EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment...

Page 14: ...mation visit www corning com lifesciences or call 800 492 1110 Outside the United States call 1 978 442 2200 or contact your local Corning sales office Corning Incorporated Life Sciences 836 North St...

Reviews: