2
| LABNET ENDURO POWER SUPPLIES
3.1 Upon Receiving the Instrument
Examine the unit carefully for any damage incurred during transit. Any damage claims must
be filed with the carrier, keep the supplied box for inspection. The warranty does not cover
in-transit damage.
To ensure safe, reliable operation, always operate the Labnet Enduro™ Power Supply in
accordance with the manufacturer’s instructions. Always wear protective gloves and safety
glasses when working in a laboratory environment. See Safety Information and Warranty
Information in this manual.
4.0 Specifications
250V
300V
Input power (switchable)
115 VAC, 50/60 Hz
230 VAC, 50/60 Hz
115 VAC, 50/60 Hz
230 VAC, 50/60 Hz
Fuse
4A/250V
2A/250V
Output power
300W
90W
Output voltage range
5 to 250V
2 to 300V
Output current range
10 to 3000 mA
4 to 500 mA
Operating modes
Constant voltage
Constant current
Constant power
1V step
10 mA step
1W step
1V step
1 mA step
–
Programmable
Yes
Store file no.: 20
Program: up to 10 steps
No
–
–
250V and 300V
Duration Timer
Approx. 99.99 hr./min.
Terminal pairs
4 positive voltage and 4 negative voltage
Crossover
Automatic
Display type
Backlit LCD graphic type
Display size (W x H)
2.11 x 0.62 in. (5.4 x 1.6 cm)
Pause function
Yes
Safety features
No load detection
Load change detection
Overload detection
Ground leak detection
Auto restart
Stackable
Yes
Housing material
Flame retardant ABS
Housing size (W x D x H)
7.9 x 12.4 x 3.4 in. (20 x 31.4 x 8.5 cm)
Operating temperature
0°C to 40°C
Environmental conditions
95% RH, 75 to 106 Kpa, Altitude not to exceed 2000 meters
Weight
5.4 lbs (2.45 kg)
Certifications
CE; cETLus
Warranty
3 years