9.3 Sterilization
Plastics
- legend to abbreviations
PS
polystyrene
ECTFE
ethylene/chlorotrifluoroethylene
SAN
styrene-acrylonitrile
ETFE
ethylene/tetrafluoroethylene
PMMA
polymethyl methacrylate
PTFE
polytetrafluoroethylene
PC
polycarbonate
FEP
tetrafluoroethylene/perfluoropropylene
PVC
polyvinyl chloride
PFA
tetrafluoroethylene/perfluoroalkylvinylether
POM
acetal Polyoxymethylenel
FKM
fluorocarbon rubber
PE-LD
low density polyethylene
EPDM
ethylene propylene diene
PE-HD
high density polyethylene
NR
natural rubber
PP
polypropylene
SI
silicon rubber
PMP
polymethylpentene
One can use all standard disinfectants. Centrifuges and devices are made of different materials, one should
consider their variety
.
radiation
β
radiation γ 25 kGy
C
2
H
4
O
(ethylene oxide)
formalin,
ethanol
PS
●
○
●
SAN
○
●
●
PMMA
●
○
●
PC
●
●
●
PVC
○
●
●
POM
●
●
●
PE-LD
●
●
●
PE-HD
●
●
●
PP
●
●
●
PMP
●
●
●
ECTFE, ETFE
○
●
●
PTFE
○
●
●
FEP, PFA
○
●
●
FKM
○
●
●
EPDM
○
●
●
NR
○
●
●
SI
○
●
●
● may be used ○ cannot be used
In the centrifuge, disinfectants and cleaning agents generally used in medical care should be used (e.g. Aerode-
sina-2000, Lysoformin 3000, Melseptol, Melsept SF, Sanepidex, Cutasept F).
9.3.1 Autoclaving
• Rotors, buckets and round carriers can be sterilized in autoclave with temperature 121°C during 20 min (215
kPa), unless otherwise specified in the OPTIONAL ACCESSORY.
• During sterilization (autoclaved) by means of steam one should to consider temperature resistance of individual
materials.
• Deformation of the accessories (carriers or lids made of plastic) may occur during autoclaving.
• Do not autoclave disposable materials (e.g. tubes, cyto-container).
• The life of the accessory depends on the frequency of autoclaving and use.
• Autoclaving reduce lifespan of plastic and mechanical components. PC tubes can become useless.
• Pressure in closed containers can cause plastic deformation or explosion.
• Prior to autoclaving the rotors and accessories, thoroughly wash and rinse with distilled water.
• Never exceed the permissible autoclaving temperature and time.
• If you want to keep the hermetic seals, replace the sealing rings after each autoclave.