Centrifuge user manual | 21
8.9 Errors
End of centrifuging – manual mode
Centrifuging may be stopped at the any moment via the STOP
key. The information message:
CYCLE CANCELLED
will be displayed.
End of centrifuging – normal mode
Stopping centrifuging in accordance the set time causes
generating multiton audible signals (after stopping the rotor)
and displaying the message
FINISH OF CENTRIFUGING
Additional messages
In case of power shortage while centrifuging, after repeated switch-
ing it on, the following error screen will be displayed:
SUPPLY DECAY WHILE CENTRIFUGING
After operating for 2000 hours, after every switching on the cen-
trifuge the error screen is being displayed with information about
the necessity to carry out servicing activities.
After pressing the SET key, the device proceeds to the main
screen and the device may operate.
Identified number of the installed rotor is not compatible with the
number of rotor remembered in program
Rotor is braking (only when centrifuge was switched off during
rotor running).
After pressing
SET
or
STOP
, the device returns to the main screen.
Screen messages that may occur during operation.
MESSAGE
EXPLANATION
“SPEED OF ROTOR”
“IDENTIFICATION <> 90 RPM”
SPEED OF ROTOR IDENTIFICATION <> 90 RPM
“IMBALANCE FAST STOP!”
“PLEASE REMOVE CAUSE”
“THEN RESTART”
UNBALANCE DETECTED
“NO ROTOR OR IDENTIFICATION”
“SENSOR DAMAGED!”
ERROR OF ROTOR IDENTIFICATION {LIMIT OF 6
SEC. IS OVER}
“INCORRECT ROTOR NUMBER!”
ROTOR’S ID NOT CORRECT
“WRONG DIRECTION OF ROTATION”
“OR UNKNOWN ROTOR!”
WRONG DIRECTION OF ROTATION / UNKNOWN
ROTOR
“PLEASE CLOSE THE LID”
“HAND!”
CLOSING THE LID MANUALLY
“ROTOR STOPPING!”
“Please wait...”
INITIALIZING AFTER MAINS FAILURE
WITH ROTATING ROTOR
“CYCLE’S ABORTED!”
CENTRIFUGING ENDED BECAUSE
OF PRESSING STOP
“CYCLE’S FINISHED”
CENTRIFUGING ENDED {WITHOUT ERRORS}