12 | Centrifuge user manual
Exemplary change of
TEMP
setting:
• Press
SET
(to enter edit mode).
• With ▲▼◄► keys mark
TEMP
fold (blinking).
• Press
SET
.
• Set demanded value by pressing ▲▼.
• Confirm set value by pressing
SET
.
• Leave edit mode by pressing
BACK
.
Changing parameters during run
There is a possibility to change parameters: SPEED, RCF,
TIME, TEMP during centrifuging. Such modifications give in
currently running program. Modification during run is repre-
sented by PROG –– symbol.
Detailed description of setting values (e.g.
TIME
).
• Press
SET
(to enter edit mode).
• With ▲▼◄► keys mark
TIME
fold (blinking).
0 0 : 0 2 : 11
[hh : mm : ss]
e.g.:
• centrifuging time – 2 minutes 11 seconds
• Press
SET
.
• Choose “hours”, “minutes” or “seconds” by pressing ◄►, e.g.:
00:07:00 (00 - blinking).
• Set demanded value by pressing ▲▼.
• Repeat above two steps to set demanded time.
• Confirm set value by pressing
SET
.
• Leave edit mode by pressing
BACK
.
set value
current value (most significant digits)
HOLD
mode
continuous operation mode
• To run centrifuging in HOLD mode set
00:00:00
time.
• To end centrifuging in HOLD mode press
STOP
.
6.5 User programs
After switching centrifuge on, program that was used in previous
session is being loaded.
Modification during run is represented by
PROG
– – symbol.
Choosing program: