6.5.
Blow-out
1.
The “Blow-out” symbol
should now be flashing.
2.
Press
«Up/Down» to select the preference (
: On,
:
Off).
The automatic blow-out function, in different pipetting modes, simulates
the blow-out function in manual pipettes.
To avoid the leftover final droplet
in tips, the blow-out setting is recommended in all protocols.
The blow-out
setting will not be performed in “Reverse Pipetting” setting.
3.
If
is selected, pressing
«SET» will confirm the selection and
move to the “PROG” setting. If
is selected, pressing
«SET»
will confirm the selection and move to the “Reverse Pipetting” setting.
Note: It is important to withdraw the tip quickly from the dispensing sample
vessel after dispense if Blow-out is selected, because the piston will retract
a short distance for creating the blow-out volume.
6.6.
Reverse Pipetting
1.
The “Reverse Pipetting” symbol
should now be flashing.
2.
Press
«Up/Down» to choose the preference (
: On,
:
Off).
The automatic reverse pipetting, in different pipetting modes, is to reserve
the final drop of sample. The BioPette E will not perform the reversed
pipetting in MIX or SE modes.
3.
Press
«SET» to confirm the selection and move to the “PROG”
setting.