ATEX and IECEx
Ex classification
Special terms (X)
Exi connection values
Maximum values in explosion group IIB
See system diagram Appendix A
VTT 16 ATEX 018X
IECEx VTT 16.0005X
II (1) G [Ex ia Ga] IIB
Ta = -30 °C…+60 °C
Uo = 14.5 V, Io = 78 mA, Po = 367 mW,
R = 243 Ω
Output voltage's characteristic curve is
trapezoidal
Co = 4.0 µF, Lo = 16.7 mH
Manufacturing year:
Please see the serial number on the type
plate.
xxxxxxx x xxx xx YY x
where YY = manufacturing year
(e.g. 18 = 2018)
9.2.
Technical specifications idOil-LIQ high liquid level sensor
TECHNICAL SPECIFICATIONS idOil-LIQ
Operating principle
Vibration
Enclosure
Enclosure class: IP 68
Materials: PVC, aluminium, PA, CR,
Viton
Weight
300 g (incl. 5 m cable)
Temparature
Operation: 0ºC…+60ºC
Environment: -30ºC…+60ºC
Supply voltage
approx. 12 V DC
Cable
2 x 0.75 mm
2
PUR
EMC
IEC/EN 61000-6-2
IEC/EN 61000-6-3
ATEX and IECEx
Ex classification
Special terms (X)
Exi connection values
Maximum values in explosion group IIB
See system diagram Appendix A
VTT 17 ATEX 005X
IECEx VTT 17.0003X
II 1 G Ex and IIB T5 Ga
Ta = -25°C…+60°C
Ui = 16 V, Ii = 80 mA, Po = 400 mW
Ci < 5.2 nF, Li < 1.6 mH
Manufacturing year:
Please see the serial number on the type
plate.
xxxxxxx x xxx xx YY x
where YY = manufacturing year
(e.g. 18 = 2018)
Installation and Operating Instructions | idOil-D30
46/51
DOC001602-EN-4