32 EN
PL-1000 Mobil
TRANSPORT AND STORAGE CONDITIONS
Controller, handpiece with handpiece cable, power supply and foot
switch
ATTENTION
Damage due to dropping
If the controller or the handpiece is dropped it can be damaged.
E
Always set the controller down on the rubber feet on a flat and stable surface so that it
cannot fall.
E
Place the handpiece back in the handpiece tray after each use so that it is safe.
E
If the controller or the handpiece is dropped, carry out a visual check of the compo-
nents.
E
Hand the controller or the handpiece in to an authorised dealer if the components
have visible damage or if they no longer function normally.
E
Transport the device and its accessories only in the original packaging.
E
Store the device and its accessories only in the following conditions:
Ambient temperature
-10°C to +40°C
Relative humidity
30% to 75%
Air pressure
700 hPa to 1070 hPa
Needle cartridge
E
Always store the needle cartridges in the following conditions:
Ambient temperature
+15°C to +35°C
Relative humidity
30% to 65%
8
8.1
8.2