Cleaning and disinfecting the equipment
ATTENTION
Damage due to fluids
If cleaning products or disinfectant gets inside the handpiece, the controller or the power
supply, this may cause a short-circuit. In addition, this may also corrode the electrical and
mechanical components.
E
Never immerse the handpiece, the controller or the power supply in cleaning product
or disinfectant.
E
Never clean the handpiece, the controller or the power supply in autoclaves or in an ul-
trasonic cleaning bath.
E
Consider material compatibilities when selecting cleaning materials and disinfectants
(see chapter "7.2 Material compatibilities" on page 30).
ATTENTION
Device damage caused by unapproved cleaning materials or disinfectants
Cleaning materials or disinfectants that are not compatible with the materials of the de-
vice can cause damage to its surface.
E
Use only cleaning materials and disinfectants that are authorised for use by the legis-
lation in your country.
E
Consider material compatibilities when selecting cleaning materials and disinfectants
(see chapter "7.2 Material compatibilities" on page 30).
E
Disconnect the controller from the power supply.
E
Wipe the controller, the handpiece cable, the handpiece and the handpiece tray with a
soft cloth moistened with cleaning product.
E
Wipe the handpiece, the controller, the handpiece cable and the handpiece tray with a
soft cloth moistened with disinfectant.